This study was undertaken to introduce a fast, accurate, selective, simple and environment-friendly colorimetric method to determine iron (II) concentration in different lipstick brands imported or manufactured locally in Baghdad, Iraq. The samples were collected from 500-Iraqi dinars stores to establish routine tests using the spectrophotometric method and compared with a new microfluidic paper-based analytical device (µPAD) platform as an alternative to cost-effective conventional instrumentation such as Atomic Absorption Spectroscopy (AAS). This method depends on the reaction between iron (II) with iron(II) selective chelator 1, 10-phenanthroline(phen) in the presence of reducing agent hydroxylamine (HOA) and sodium acetate (NaOAc) b
... Show MoreThe adsorption ability of Iraqi initiated calcined granulated montmorillonite to adsorb Symmetrical Schiff Base Ligand 4,4’-[hydrazine-1, 2-diylidenebis (methan-1-yl-1-ylidene)) bis (2-methoxyphenol)] derived from condensation reaction of hydrazine hydrate and 4-hydroxy-3-methoxybenzaldehyde, from aqueous solutions has been investigated through columnar method.The ligand (H2L) adsorption found to be dependent on adsorbent dosage, initial concentration and contact time.All columnar experiments were carried out at three different pH values (5.5, 7and 8) using buffer solutions at flow rate of (3 drops/ min.),at room temperature (25±2)°C. The experimental isotherm data were analyzed using Langmuir, Freundlich and Temkin equations. The monol
... Show MoreThe current research aims to show the impact of the international auditing standard IAS 540 in reducing income smoothing practices in Iraq. To achieve the objectives of the research, the researcher adopted a questionnaire for a sample of auditors in Iraq. Where 60 forms were distributed and after the questionnaire was retrieved and statistical analysis was done using the SPSS program, The research reached a number of results, the most important of which are: the existence of a statistically significant effect of the application of the international auditing standard IAS 540 in reducing income smoothing practices, The research recommended the necess
... Show MoreA new method is characterized by simplicity, accuracy and speed for determination of Oxonuim ion in ionisable inorganic acid such as hydrochloric (0.1 - 10) ,Sulphuric ( 0.1 - 6 ),nitric ( 0.1 - 10 ), perchloric ( 0.1 - 7 ), acetic (0.1 - 100 ) and phosphoric ( 0.1 - 30 ) ( mMol.L-1 )acids. By continuous flow injection analysis. The proposed method was based on generation of bromine from the Bro-3-Br-- H3O+. Bromine reacts with fluorescein to quenches the fluorescence . A sample volume no.1 (31μl) and no.2 (35μl) were used with flow rate of 0.95 mL.min-1 using H2O line no.1as carrier stream and 1.3 mL.min-1 using fluorescein sodium salt line no.2. Linear regression of the concentration ( mMol.L-1 ) Vs quenched fluorescence gives a correla
... Show MoreCo(II) ion was determined by a new, accurate, sensitive and rapid method via a
continuous flow injection analysis (CFIA) with a chemiluminescence reaction based on
the oxidation of Luminol which is loaded on poly acrylic acid gel beads by hydrogen
peroxide in presence of Cobalt (II) ion as a chemiluminescence catalyst. Chemical and
physical parameters were investigated to obtain the best conditions. Linear dynamic
range of Cobalt (II) ion was from 0.1-20.0 μg.ml-1 with a correlation coefficient r =
0.9758, limit of detection (L.O.D) 0.2 ng/sample from the step wise dilution of lowest
concentration in the calibration graph with the percentage relative standard deviation for
3 μg.ml-1 Co(ll) solution is 0.8537% (n
The study aimed at identifying the role of the Faculty of Education in the community service in light of vision of the Kingdom 2030 from the faculty members’ point of view at the University of Jeddah. The study used a descriptive method through a questionnaire administered to a sample of the faculty members.It concluded that the role of the College in serving the community is the development of the right habits in the family and the importance of establishing bridges of cooperation between the Faculty of Education and the surrounding environment and teaching special education teachers how to apply some methods of measurement to determine the needs of students with disabilities. The results also pointed out that there are no statistical
... Show MoreA new simple sensitive and selective spectrophotometric method has been
developed for the analysis of vanadium(V) in three randomly chosen samples from
river water at different locations by continuous flow injection analysis. The method
based on the oxidation of pyrogallol by vanadium(V) in acidic solution to form color
species and the same species was determined using homemade Ayah 6SX1-T-2D
solar cell analyser . Chemical and physical parameters were investigated using the
high intensity of snow white light emitted diode as a source. The linear dynamic
range for the instrument response versus vanadium(V) concentration was 1-200
mg.L-1 with correlation coefficient r = 0.9920. The limit of detection (S/N=3) was 70<
This piece of research work aims to study one of the most difficult reaction and determination due to continuous and rapid variation of reaction products and the reactants. As molybdenum (VI) aid in the decomposition of hydrogen peroxide in alkaline medium of ammomia, thus means a continuous liberation of oxygen which cuases and in a continuous manner a distraction in the measurement process. On this basis pyrogallol was used to absorbe all liberated oxygen and the result is an a clean undisturbed signals. Molybdenum (VI) was determined in the range of 4-100 ?g.ml-1 with percentage linearity of 99.8% or (4-300 ?g.ml-1 with 94.4%) while L.O.D. was 3.5 ?g.ml-1. Interferring ions (cations and anions) were studied and their main effect was red
... Show MoreA newly developed analytical method characterized by its speed and sensitivity for
the determination of cadmium (II) in aqueous solution in three randomly chosen
samples from river water at different locations via turbidimetric measurement by
Ayah 6SX1-T-2D Solar - CFI analyser. The method is based on the formation of
yellowish white precipitate for the complex Cd3[Fe(CN)6]2 by direct reaction of the
cadmium (II) with potassium hexacyano ferrate (III) in aqueous medium. Turbidity
was measured via the reflection of incident light that collides on the surfaces
precipitated particles at 0-180o. Chemical and physical parameters were investigated.
Linear dynamic of cadmium (II) is ranged from 0.05-12 mmol.L-1, with cor