Preferred Language
Articles
/
ijs-10594
Chemiluminometric- CFIA for the Determination of Cobalt (ll) Ion in Commercial Cobalt-Molybdenum Catalyst (K F124-3E) Used in Desulphurization Processes of Petroleum Products via Multi Gel Beads Reactor.
...Show More Authors

Co(II) ion was determined by a new, accurate, sensitive and rapid method via a
continuous flow injection analysis (CFIA) with a chemiluminescence reaction based on
the oxidation of Luminol which is loaded on poly acrylic acid gel beads by hydrogen
peroxide in presence of Cobalt (II) ion as a chemiluminescence catalyst. Chemical and
physical parameters were investigated to obtain the best conditions. Linear dynamic
range of Cobalt (II) ion was from 0.1-20.0 μg.ml-1 with a correlation coefficient r =
0.9758, limit of detection (L.O.D) 0.2 ng/sample from the step wise dilution of lowest
concentration in the calibration graph with the percentage relative standard deviation for
3 μg.ml-1 Co(ll) solution is 0.8537% (n=5). The method was applied successfully for the
determination of Co (II) ion in CoMo catalyst type ( K F124- 3E).

View Publication Preview PDF
Quick Preview PDF
Publication Date
Sun Aug 13 2017
Journal Name
Ibn Al-haitham Journal For Pure And Applied Sciences
Separation of cobalt from spent catalyst CoMo by precipitation and Ion exchange techniques
...Show More Authors

Two methods were established to separate cobalt from the spent catalyst CoMo which also contain Co, Al and Fe. The first method was the precipitation technique by controlling the pH. At pH 5, 76% of the cobalt which was collected with 1.4% Al and 0.5% Fe as contaminants. The second method was the anion exchange by using Amberlite 400 resin, 100% of the cobalt and was collected with 99.46% purity.The only contaminant was Fe with 0.54% with no Al.  For a large scale production of cobalt from this spent catalyst, a batch process was designed with a production of 80 grams per batch by using the anion exchange technique. Kilograms quantities of Co were collected.

View Publication Preview PDF
Publication Date
Thu Dec 01 2022
Journal Name
Baghdad Science Journal
Determination of Nickel and Cobalt in Cosmetic Products Marketed in Iraq Using Spectrophotometric and Microfluidic Paper-based Analytical Device (µPADs) Platform
...Show More Authors

Two quantitative, environment-friendly and easily monitored assays for Ni (II) and Co (III) ions analysis in different lipstick samples collected from 500-Iraqi dinars stores located in Baghdad were introduced. The study was based on the reaction of nickel (II) ions with dimethylglyoxime (DMG) reagent and the reaction of cobalt (III) ions with 1-nitroso-2-naphthol (NN) reagent to produce colored products. The color change was measured by spectrophotometric method at 565 nm and 430 nm for Ni and Co, respectively, with linear calibration graphs in the concentration range 0.25-100 mg L-1 (Ni) and 0.5-100 mg L-1 (Co) and LOD and LOQ of 0.11 mg L-1 and 0.36 mg L-1 (Ni), and 0.15 mg L-1 an

... Show More
View Publication Preview PDF
Scopus (4)
Crossref (3)
Scopus Clarivate Crossref
Publication Date
Sat Dec 30 2023
Journal Name
Iraqi Journal Of Science
Continuous Flow Injection Analysis for The Photometric Determination of Metformin Drug Via The Release of Copper(II) ion from Charged Gel Bead Crystal
...Show More Authors

A newly developed analytical method characterized by its speed and sensitivity for the determination of metformin-HCl via the formation of complex for metformin-HCl-OH--copper(II) ion from the gel bead system by continuous flow injection analysis. The method is based on the imbedded copper(II) ion in the gel bead structure can be used in the reaction for the formation of red –magneta colour complex(λmax. =530nm) formed by direct reaction of the drug with the released copper (II) ion from the gel bead in alkaline medium. Linear dynamic range for the absorbance versus metformin concentration was 0.001-1 mmol.L-1 while C.O.D was (r2% =95.33%) . The L.O.Q was 0.868mmol.L-1.L.O.D (S/N=3)=0.5μmol.L-1 from the step wise dilution for the min

... Show More
View Publication Preview PDF
Publication Date
Thu Nov 30 2023
Journal Name
Iraqi Journal Of Science
CFIA-ISNAG fluorimeter for the Determination of Bromhexine-HCl in Drugs via the Measurement Scattered Light at ± 90˚
...Show More Authors

A precise, simple, and accurate continuous flow injection technique was used for the instantaneous estimation of bromhexine hydrochloride (BH-HCl) in tablet dosage form. The chemical and physical parameters of the reaction of BH-HCl with tetraphenylborate to produce a yellowish-white precipitate were determined using an ISNAG fluorimeter analyzer and diverging light at 90°. The calibration curve for BH-HCl was linear with correlation coefficients of 0.9994 and linearity percentage r2% = 99.87 over a concentration range of 0.01-20 mmol/L, L.O.D = 0.3610 µg/125µL (0.007 mmol/L), and RSD% less than 0.3% for 3 and 13 mmol/L (five replicates). This approach was efficiently used to estimate the levels of BH-HCl in two distinct pharmaceutica

... Show More
View Publication Preview PDF
Scopus Crossref
Publication Date
Thu Mar 30 2017
Journal Name
Iraqi Journal Of Pharmaceutical Sciences ( P-issn 1683 - 3597 E-issn 2521 - 3512)
Spectroscopic Study for Determination of Amoxicillin Using Cobalt(II) as Complexing Metal
...Show More Authors

         This study includes analytical methods for the determination of the drug amoxicillin trihydrate (Amox.) in some pharmaceutical preparations using Cobalt ion (Co(II)) as complexing metal. The best conditions for complexation were: the reaction time was 20 minutes, pH=1.5 and the best temperature of reaction was 70 ËšC. Benzyl alcohol was the best solvent for extraction the complex.

Keywords: Amoxicillin, Cobalt(II), Complex, Molar ratio.

View Publication Preview PDF
Crossref
Publication Date
Sat Jun 03 2023
Journal Name
Iraqi Journal Of Science
New mode for the on-line determination of oxonium ion in different strong acids using CFIA via the use of homemade linear array Ayah 5Sx4-ST- 5D solar CFI analyser
...Show More Authors

New mode for the on-line determination of oxonium ion in different strong acids using CFIA via the use of homemade linear array Ayah 5Sx4-ST- 5D solar CFI analyser

View Publication Preview PDF
Publication Date
Sun Apr 30 2023
Journal Name
Iraqi Journal Of Science
New Mode of On- Line Automation: Chemilumino-Fluoro Metric Method for the Determination of Hydrogen Peroxide by Fluorescence Energy Transfer Using Ploy Acrylic Acid Gel Beads
...Show More Authors

Hydrogen peroxide was determined by a new , accurate , sensitive and rapid method via continuous mode of FIA coupled with total luminescence measurement which include the chemiluminescence generated ,based on the oxidation of Luminol which is loaded on poly acrylic acid gel beads by hydrogen peroxide in presence of Cobalt (II) ion as a chemiluminescence catalyst and the fluorescence that was created by the insitu radiation of the released chemiluminescence light. Fluorescien molecule was used as an accepter fluorophore where it is irradiated internally and instantly by the generation of luminol chemiluminescence light as internal source for irradiation of fluorescien molecule (Fluorescence Energy Transfer (FRET) ) . It can easily give fl

... Show More
View Publication Preview PDF
Publication Date
Fri Jan 26 2024
Journal Name
Iraqi Journal Of Science
Histological Changes in the Duodenum of Mice Treated with Cobalt
...Show More Authors

The present study was conducted using 16 adult male Swiss albino mice (weighing 40-50g). They were equally divided into four experimental groups. The first group was only given tap water as control during experimentation period (40 days). Group II, group III and group IV of mice were daily given a subcutaneous injection of cobalt as cobalt chloride (CoCl2) at a dose of 20mg, 40mg, and 60mg /Kg body weight respectively. The examination of the histological sections of the second experimental group of mice (low dose group) showed no particular changes in the columnar epithelial cells of the duodenum of mice. The examination of the histological sections of the third experimental group of mice (moderate dose group) showed abundant degenerativ

... Show More
View Publication Preview PDF
Publication Date
Thu Jun 10 2021
Journal Name
Journal Of Kufa−physics
The Structural and Optical Properties of Cobalt dioxide (CoO2 )Thin Films deposited via (SCSP) Technique for photovoltaic applications
...Show More Authors

Publication Date
Thu Oct 01 2020
Journal Name
Chemical Engineering And Processing - Process Intensification
Cobalt Removal from Simulated Wastewaters Using a Novel Flow-by Fixed Bed Bio-electrochemical Reactor
...Show More Authors

View Publication
Scopus (28)
Crossref (21)
Scopus Clarivate Crossref