...Show More Authors
This study was done in Baghdad teaching Hospital by using developed instrument type GIOHO and included a number of patients with compressed breast thickness (7,8,9,10)cm .
The relationship between radiation dose and breast thickness was linear. All results were compared with the international standered values that measured by the International Nuctear Agency and Europeon sources ,it was found that it is in consistance or has a little difference .
The study showed that the mean absorbed dose may be determined by using TLD measurement below 10 mGy and the glandular dose was (1.45 mGy) and this can not b
...
Show More