A new simple sensitive and selective spectrophotometric method has been developed for the analysis of vanadium(V) in three randomly chosen samples from river water at different locations by continuous flow injection analysis. The method based on the oxidation of pyrogallol by vanadium(V) in acidic solution to form color species and the same species was determined using homemade Ayah 6SX1-T-2D solar cell analyser . Chemical and physical parameters were investigated using the high intensity of snow white light emitted diode as a source. The linear dynamic range for the instrument response versus vanadium(V) concentration was 1-200 mg.L-1 with correlation coefficient r = 0.9920. The limit of detection (S/N=3) was 70 ng/ sample from the step wise dilution for the minimum concentration in the linear dynamic range of the calibration graph with RSD % of lower than 1% at 90 mg.L-1 (n=5) concentration of vanadium(V). The method was applied successfully for the determination of vanadium (V) in three river samples. A comparison was made between: both of the proposed methods and classical method (UV-Vis spectrophotometry at wave length 427 nm) using the standard additions method via the use of paired t-test. It was noticed that there was no significant difference between two methods at 95 % confidence level.
A new turbidimetric-flow injection method is described for the determination of chlorpromazine HCl in pure and pharmaceutical preparation. The method is characterized by simplicity, sensitivity and fast, it is based on formation of ion pair compound between chlorpromazine HCl and Potassium hexacyanoferrate(III) in an acid medium for the formation of greenish yellow precipitate. This precipitate was determined using homemade Linear Array Ayah 5SX1-T-1D continuous flow injection analyser. Optimum concentrations of chemical reactants, physical instrumental conditions have been investigated. The linear dynamic range of chlorpromazine HCl was 3-30 mmol.L-1 while correlation coefficient (r) was 0.9929 and percentage linearity (%r2) C.O.D was 9
... Show MoreA new, simple, sensitive and fast developed method was used for the determination of methyldopa in pure and pharmaceutical formulations by using continuous flow injection analysis. This method is based on formation a burgundy color complex between methyldopa andammonium ceric (IV) nitrate in aqueous medium using long distance chasing photometer NAG-ADF-300-2. The linear range for calibration graph was 0.05-8.3 mmol/L for cell A and 0.1-8.5 mmol/L for cell B, and LOD 952.8000 ng /200 µL for cell A and 3.3348 µg /200 µL for cell B respectively with correlation coefficient (r) 0.9994 for cell A and 0.9991 for cell B, RSD % was lower than 1 % for n=8. The results were compared with classical method UV-Spectrophotometric at λ max=280 n
... Show MoreThe proposed method is sensitive, simple , fast for the determination of mebeverine hydrochloride in pure form or in pharmaceutical dosage . Using Homemade instrument fluorimeter continuous flow injection analyser with solid state laser (405 nm) as a source. Where it is based upon the fluorescence of fluorescein sodium salt and quenching effect of fluorescence by mebeverine in aqueous medium. The calibration graph was linear in the concentration range 0.05 to10 mMol.L-1 (r= 0.9629) with relative standard deviation (RSD%) for 1 mMol.L-1mebeverine solution was lower than 3% (n=6). Three pharmaceutical drugs were used as an application for the determination of mebeverine. A comparison was made between the newly developed method of analysis wit
... Show MoreThe objective of this study was to establish an accurate, precise, sensitive, simple, fast and reliable method for the determination of ciprofloxacin hydrochloride in pure or in pharmaceutical dosage forms using Homemade instrument fluorimeter continuous flow injection analyser with solid state laser (405 nm) as a source. The method is based upon the fluorescence fluorescein sodium salt and its quenching by ciprofloxacin hydrochloride in aqueous medium. The solutions of standard and the sample were prepared in distilled water. The calibration graph was linear in the concentration range of using (10 - 100) mMol.L-1 ciprofloxacin hydrochloride (r= 0.9891) with relative standard deviation (RSD%) for 3 mMol.L-1 ciprofloxacin HCl solution is
... Show MoreA newly photometric analytical method characterized by its speed and sensitivity was developed for the determination of folic acid in pure and pharmaceutical samples via its oxidation to reddish-orange coloured complex through oxidation by cerium (IV) sulphate in aqua medium using homemade Ayah 3Sx3-3D-solar cell CFI photometer. The colored species were determined using supper bright green light emitting diode (LED) as a source. A 100μl was taken as a best sample volume for the determination of folic acid. The linearity of calibration curve for the instrument response versus folic acid concentration was 0.005-20 mmol.L-1 while the L.O.D. was 0.5 mol.L-1 from the stepwise dilution for the minimum concentration of lowest concentration
... Show MoreA new method is characterized by simplicity, accuracy and speed for determination of Oxonuim ion in ionisable inorganic acid such as hydrochloric (0.1 - 10) ,Sulphuric ( 0.1 - 6 ),nitric ( 0.1 - 10 ), perchloric ( 0.1 - 7 ), acetic (0.1 - 100 ) and phosphoric ( 0.1 - 30 ) ( mMol.L-1 )acids. By continuous flow injection analysis. The proposed method was based on generation of bromine from the Bro-3-Br-- H3O+. Bromine reacts with fluorescein to quenches the fluorescence . A sample volume no.1 (31μl) and no.2 (35μl) were used with flow rate of 0.95 mL.min-1 using H2O line no.1as carrier stream and 1.3 mL.min-1 using fluorescein sodium salt line no.2. Linear regression of the concentration ( mMol.L-1 ) Vs quenched fluorescence gives a correla
... Show MoreA new, simple, accurate, fast and sensitive spectrophotometric method has been developed for the analysis of amiloride in pure and pharmaceutical formulations by continuous flow injection analysis. The method was based on the oxidation of the drug with Ce(IV)sulfate in acidic medium to formed a pinkish-red color species which determined using homemade Ayah 3SBGR x3-3D solar cell flow injection photometer. Optimum conditions were obtained using a high intensity green light emitted diode as a source. The linear dynamic range for the instrument response versus amiloride concentration was 0.0005- 10 mmol.L-1 while the L.O.D was of 9.471 ng / sample. The correlation coefficient (r) was 0.9824 while percentage linearity (r2%) was 96.50%. RSD%
... Show MoreA newly developed analytical method was conducted for the determination of Ketotifen fumarate (KTF) in pharmaceuticals drugs via quenching of continuous fluorescence of 9(10H)-Acridone (ACD). The method was applied using flow injection system of a new homemade ISNAG fluorimeter with fluorescence measurements at ± 90◦ via 2×4 solar cell. The calibration graph was linear in the range of 1-45 mmol/L, with correlation coefficient r = 0.9762 and the limit of detection 29.785 µg/sample from the stepwise dilution for the minimum concentration in the linear dynamic ranged of the calibration graph. The method was successfully applied to the determination of Ketotifen fumarate in two different pharma
... Show MoreThe research work represent a fast and simple method for the determination of methionine using chemiluminescence for the methionine-sodium hydroxide-luminol for the generation of a chemiluminesecent derivative of luminal. The emission was measured by continuous flow analysis made sample size of 83µL was used.Response versus concentration extended from 0.2-20 mM.L-1 with a percentage linearity of 96.17% or with 99.17% percentage of linearity for the range 0.6-20 mM.L-1. Reaching to a L.O.D. at (S/N=3) for 5 µM.L-1 from the gradual dilution for the minimum concentration in the calibration graph with a repeatability of less than 0.5% (n=10). A comparison was made between the new developed method with the classical method for the spectrophoto
... Show More