Preferred Language
Articles
/
ijs-11739
New Turbidimetric-Continuous Flow Injection Analysis Method for The Determination of Chlorpromazine HCl in Pharmaceutical Preparation Using Linear Array Ayah 5SX1-T-1D-CFI Analyser
...Show More Authors

A new turbidimetric-flow injection method is described for the determination of chlorpromazine HCl in pure and pharmaceutical preparation. The method is characterized by simplicity, sensitivity and fast, it is based on formation of ion pair compound between chlorpromazine HCl and Potassium hexacyanoferrate(III) in an acid medium for the formation of greenish yellow precipitate. This precipitate was determined using homemade Linear Array Ayah 5SX1-T-1D continuous flow injection analyser. Optimum concentrations of chemical reactants, physical instrumental conditions have been investigated. The linear dynamic range of chlorpromazine HCl was 3-30 mmol.L-1 while correlation coefficient (r) was 0.9929 and percentage linearity (%r2) C.O.D was 98.59%. Limit of Detection (S/N=3) 3×10-7 M/sample equivalent to 0.12 g/sample from the stepwise dilution of minimum concentration for the lowest concentration in the linear dynamic range of the calibration graph with R.S.D.% (n=6) < 2% for concentration 8 and 10 mmol.L-1. The method was applied successfully for the determination of chlorpromazine HCl in pharmaceutical drugs. A comparison was made between the developed method with the official method via the use of paired t-test. It shows that there were no significant differences between either methods. Therefore the newly developed method (K3[Fe(CN)6]-HCl-Chlorpromazine HCl) can be adopted as an alternative method for determination of Chlorpromazine HCl.

View Publication Preview PDF
Quick Preview PDF
Publication Date
Sun Jul 02 2023
Journal Name
Iraqi Journal Of Science
A New Approach for the Turbidimetric Determination of Hydronium ion by Using Homemade Linear Array Ayah 5SX1-T-1D-CFI Analyser
...Show More Authors

A new Turbidimetric method characterized by simplicity, accuracy and speed for determination of Hydronium ion by continuous flow injection analysis. The method was based on the formation of complex Zn3[Fe(CN)6] for Zinc(II) that was eluted by Hydronium ion from cation exchanger column with Potassium hexacyanoferrate(III) for the formation of a pale yellow precipitate and this precipitate was determined using homemade Linear Array Ayah-5SX1-T-1D continuous flow injection analyser. The optimum parameters were 2.7 mL.min-1 flow rate using H2O as a carrier stream, 1.7 mL.min-1 reagent stream, 110 L sample volume and open valve for the purge of the sample segment. Data treatment shows that linear range 0.01-0.1 mol.L-1 for each acids (HClO

... Show More
View Publication Preview PDF
Publication Date
Sun Oct 22 2023
Journal Name
Iraqi Journal Of Science
A New Approach for the Turbidimetric Determination of Hydronium ion by Using Homemade Linear Array Ayah 5SX1-T-1D-CFI Analyser
...Show More Authors

A new Turbidimetric method characterized by simplicity, accuracy and speed for determination of Hydronium ion by continuous flow injection analysis. The method was based on the formation of complex Zn3[Fe(CN)6] for Zinc(II) that was eluted by Hydronium ion from cation exchanger column with Potassium hexacyanoferrate(III) for the formation of a pale yellow precipitate and this precipitate was determined using homemade Linear Array Ayah-5SX1-T-1D continuous flow injection analyser. The optimum parameters were 2.7 mL.min-1 flow rate using H2O as a carrier stream, 1.7 mL.min-1 reagent stream, 110 L sample volume and open valve for the purge of the sample segment. Data treatment shows that linear range 0.01-0.1 mol.L-1 for each acids (HClO

... Show More
View Publication Preview PDF
Publication Date
Sun Oct 22 2023
Journal Name
Iraqi Journal Of Science
New Approach for The Turbidimetric Determination of Iron(III) in Drug Samples of Different Origins Using Homemade Linear Array Ayah 5SX1-T-1D-CFI Analyser
...Show More Authors

A new Turbidimetric method characterized by simplicity, accuracy and speed for determination of iron(III) in drug samples by continuous flow injection analysis. The method was based on the formation of complex for iron(III) with 8-hydroxyquinoline in presence of ammonium acetate as a medium for the formation of deep green precipitate and this precipitate was determined using homemade Linear Array Ayah-5SX1-T-1D continuous flow injection analyser. The optimum parameters were 2.6 mL.min-1 flow rate using H2O as a carrier, 1.9 mL.min-1 (14 mmol.L-1) ammonium acetate, 2.4 mL.min-1 (14 mmol.L-1) 8-hydroxyquinoline, 60 L sample volume and open valve for the purge of the sample segment. Data treatment shows that linear range 0.1-8.0 mmol.L-1

... Show More
View Publication Preview PDF
Publication Date
Sun Oct 22 2023
Journal Name
Iraqi Journal Of Science
New Developed Data Treatments for the Characteristic Linear Array Ayah 5SX1-T-1D-CFI Analyser Segment Response Profile & Generalization
...Show More Authors

Derivative spectrophotometry is one of the analytical chemistry techniques used
in the analysis and determination of chemicals and pharmaceuticals. This method is
characterized by simplicity, sensitivity and speed. Derivative of Spectra conducted
in several ways, including optical, electronic and mathematical. This operation
usually be done within spectrophotometer. The paper is based on form of a new
program. The program construction is written in Visual Basic language within
Microsoft Excel. The program is able to transform the first, second, third and fourth
derivatives of data and the return of these derivatives to zero order (normal plot).
The program was applied on experimental (trial) and reals values of su

... Show More
View Publication Preview PDF
Publication Date
Sun Apr 30 2023
Journal Name
Iraqi Journal Of Science
Turbidimetric Determination of Metoclopramide Hydrochloride in Pharmaceutical Preparation via the Use of A new Homemade Ayah 6SX1-T-2D Solar Cell-Continuous Flow Injection Analyser
...Show More Authors

A newly developed analytical method characterized by its speed and sensitivity for the determination of metoclopramide hydrochloride (MCP-HCl) in pure and pharmaceutical preparation via turbidimetric measurement (0-180o) by Ayah 6SX1-T-2D Solar cell-CFI Analyser. The method was based on the reaction of phosphomolybdic acid with metoclopramide hydrochloride in acidic medium to form yellowish white precipitate for the ion-pair complex. Turbidity was measured via the reflection of incident light that collides on the surface precipitated particles at 0-180o. Chemical and physical parameters were studied and optimized. The calibration graph was linear in the range of 0.0005-3 or 0.0005- 4 mMol.L-1, with correlation coefficient r = 0.9947&

... Show More
View Publication Preview PDF
Publication Date
Sun Oct 22 2023
Journal Name
Iraqi Journal Of Science
New Mode for the On-Line Determination of Amiloride in Pure and Pharmaceutical Preparation Using CFIA Via the Use of Linear Array from Six LED (Snow White) with One Solar Cell in a Homemade Ayah 6Sx1-T-1D Solar CFI Analyser
...Show More Authors

A newly flow injection-turbidimetric method characterized by it is speed and sensitivity has been developed for the determination of Amiloride in pure and pharmaceutical preparations. It is based on the formation of yellowish white precipitate for the Amiloride-phosphomolybidic acid ion pair in aqueous medium. Turbidity was measured by Ayah 6Sx1-T-1D solar cell CFI analyser via the attenuation of incident light from the surfaces precipitated particles at 0-180. The Chemical and physical parameters were investigated. Linear dynamic range for the attenuation of incident light versus Amiloride concentration was of 0.005-10 mmol.L-1, with the correlation coefficient (r) of 0.9986 , while the percentage linearity (r2%) was 99.71%. The L.O.

... Show More
View Publication Preview PDF
Publication Date
Sun Jul 02 2023
Journal Name
Iraqi Journal Of Science
New Mode for the On-Line Determination of Amiloride in Pure and Pharmaceutical Preparation Using CFIA Via the Use of Linear Array from Six LED (Snow White) with One Solar Cell in a Homemade Ayah 6Sx1-T-1D Solar CFI Analyser
...Show More Authors

A newly flow injection-turbidimetric method characterized by it is speed and sensitivity has been developed for the determination of Amiloride in pure and pharmaceutical preparations. It is based on the formation of yellowish white precipitate for the Amiloride-phosphomolybidic acid ion pair in aqueous medium. Turbidity was measured by Ayah 6Sx1-T-1D solar cell CFI analyser via the attenuation of incident light from the surfaces precipitated particles at 0-180. The Chemical and physical parameters were investigated. Linear dynamic range for the attenuation of incident light versus Amiloride concentration was of 0.005-10 mmol.L-1, with the correlation coefficient (r) of 0.9986 , while the percentage linearity (r2%) was 99.71%. The L.O.

... Show More
View Publication Preview PDF
Publication Date
Sun Oct 22 2023
Journal Name
Iraqi Journal Of Science
New Approach for an On-Line Turbidimetric Determination of Cadmium (II) in Different River Water Via the Use of A homemade Ayah 6SX1-T-2D Solar-Continuous Flow Injection Analyser
...Show More Authors

A newly developed analytical method characterized by its speed and sensitivity for
the determination of cadmium (II) in aqueous solution in three randomly chosen
samples from river water at different locations via turbidimetric measurement by
Ayah 6SX1-T-2D Solar - CFI analyser. The method is based on the formation of
yellowish white precipitate for the complex Cd3[Fe(CN)6]2 by direct reaction of the
cadmium (II) with potassium hexacyano ferrate (III) in aqueous medium. Turbidity
was measured via the reflection of incident light that collides on the surfaces
precipitated particles at 0-180o. Chemical and physical parameters were investigated.
Linear dynamic of cadmium (II) is ranged from 0.05-12 mmol.L-1, with cor

... Show More
View Publication Preview PDF
Publication Date
Fri Dec 08 2023
Journal Name
Iraqi Journal Of Science
Determination of Atenolol in pharmaceutical formulations by continuous flow injection analysis via turbidimetric (T180o) and scattered light effect at two opposite position (2N90o ) using Ayah 4SW-3D-T180 -2N90 -Solar - CFI
...Show More Authors

A simple, rapid and sensitive method for the analysis of Atenolol in pure and pharmaceutical preparation as an alternative analytical procedure were developed by continuous flow injection analysis via turbidimetric (T180o) and scattered light effect at two opposite position (2N90o). The method is based upon the formation of white precipitate for the ion pair compound by phosphomolybidic acid with Atenolol in aqueous medium. The precipitate is measured via the attenuation of incident light and scattering of the incident light in two opposite direction namely +90o and -90o angle were measured. Chemical and physical parameters were investigated. The linearity of Atenolol is ranged from (0.1-11) mmol.L-1, with correlation coefficient r=0.993

... Show More
View Publication Preview PDF
Publication Date
Sun Dec 01 2013
Journal Name
Baghdad Science Journal
Flow injection analysis for the photometric determination of promethazine-HCl in pure and pharmaceutical preparation via oxidation by persulphate using Ayah 3SX3-3D solar micro photometer
...Show More Authors

The first flow injection spectrophotometric method is characterized by its speed and sensitivity which have been developed for the determination of promethazine-HCl in pure and pharmaceutical preparation. It is based on the in situ detection of colored cationic radicals formed via oxidation of the drug with sodium persulphate to pinkish-red species and the same species was determined by using homemade Ayah 3SX3-3D solar flow injection photometer. Optimum conditions were obtained by using the high intensive green light emitted diode as a source. Linear dynamic range for the absorbance versus promethazine-HCl concentration was 0-7 mmol.L-1, with the correlation coefficient (r) was 0.9904 while the percentage linearity (r2%) was 98.09%. the L.

... Show More
View Publication Preview PDF
Crossref