Determination of Methionine Via Chemiluminescence-Continuous Flow Injection Analysis
...Show More Authors

The research work represent a fast and simple method for the determination of methionine using chemiluminescence for the methionine-sodium hydroxide-luminol for the generation of a chemiluminesecent derivative of luminal. The emission was measured by continuous flow analysis made sample size of 83µL was used.Response versus concentration extended from 0.2-20 mM.L-1 with a percentage linearity of 96.17% or with 99.17% percentage of linearity for the range 0.6-20 mM.L-1. Reaching to a L.O.D. at (S/N=3) for 5 µM.L-1 from the gradual dilution for the minimum concentration in the calibration graph with a repeatability of less than 0.5% (n=10). A comparison was made between the new developed method with the classical method for the spectrophotometric determination by coupling with paired t-test or F-test and found that there is no significant difference between the two methods and the new adapted method which can be used as an alternative method.

View Publication
Publication Date
Fri Dec 08 2023
Journal Name
Iraqi Journal Of Science
Determination of Atenolol in pharmaceutical formulations by continuous flow injection analysis via turbidimetric (T180o) and scattered light effect at two opposite position (2N90o ) using Ayah 4SW-3D-T180 -2N90 -Solar - CFI
...Show More Authors

A simple, rapid and sensitive method for the analysis of Atenolol in pure and pharmaceutical preparation as an alternative analytical procedure were developed by continuous flow injection analysis via turbidimetric (T180o) and scattered light effect at two opposite position (2N90o). The method is based upon the formation of white precipitate for the ion pair compound by phosphomolybidic acid with Atenolol in aqueous medium. The precipitate is measured via the attenuation of incident light and scattering of the incident light in two opposite direction namely +90o and -90o angle were measured. Chemical and physical parameters were investigated. The linearity of Atenolol is ranged from (0.1-11) mmol.L-1, with correlation coefficient r=0.993

... Show More
View Publication Preview PDF
Publication Date
Tue Oct 01 2019
Journal Name
International Journal Of Drug Delivery Technology
Simple and Rapid Method For Estimate of Propranolol With Bi (III) Via Long-Distance Chasing Photometer (NAG-ADF-300-2) Utilization Continuous Flow Injection Analysis
...Show More Authors

A simple, sensitive and rapid method was used for the estimate of: Propranolol with Bi (III) to prove the efficiency, reliability and repeatability of the long distance chasing photometer (NAG-ADF-300-2) using continuous flow injection analysis. The method is based on a reaction between propranolol and Bi (III) in an aqueous medium to obtain a yellow precipitate. Optimum parameters were studied to increase the sensitivity for the developed method. A linear range for calibration graph was 0.1-25 mmol/L for cell A and 1-40 mmol/L for cell B, and LOD 51.8698 ng/200 µL and 363.0886 ng /200 µL , respectively to cell A and cell B with correlation coefficient (r) 0.9975 for cell A, 0.9966 for cell B, RSD% was lower than 1%, (n = 8) for the

... Show More
Publication Date
Fri Jun 16 2023
Journal Name
Iraqi Journal Of Pharmaceutical Sciences ( P-issn 1683 - 3597 E-issn 2521 - 3512)
Colorimetric Determination of Salbutamol Sulfate using Spectrophotometry-Continuous Flow Injection Technique in Pure and Pharmaceutical Forms
...Show More Authors

Simple, precise and economic batch and flow injection analysis (FIA)-spectrophotometric methods have been established for simultaneous determination of salbutamol sulfate (SLB) in bulk powder and pharmaceutical forms. Both methods based on diazotization coupling reaction of SLB with another drug compound (sulfadimidine) as a safe and green diazotization agent in alkaline medium. At 444 nm, the maximum absorption of the orange azo-dye product was observed. A thorough investigation of all chemical and physical factors was conducted for batch and FIA procedures to achieve high sensitivity. Under the optimized experimental variables, SLB obeys Beer’s law in the concentration range of 0.25-4 and 10-100 μg/mL with limits of detection o

... Show More
Scopus (4)
Crossref (2)
Scopus Crossref
Publication Date
Mon Mar 23 2020
Journal Name
Baghdad Science Journal
A Assessment of long distance chasing photometer (NAG-ADF-300-2) by estimating the drug atenolol with ammonium molybdate via continuous flow injection analysis: drug atenolol a
...Show More Authors

     Atenolol was used with ammonium molybdate to prove the efficiency, reliability and repeatability of the long distance chasing photometer (NAG-ADF-300-2) using continuous flow injection analysis. The method is based on reaction between atenolol and ammonium molybdate in an aqueous medium to obtain a dark brown precipitate. Optimum parameters was studied to increase the sensitivity for developed method. A linear range for calibration graph was 0.1-3.5 mmol/L for cell A and 0.3-3.5 mmol/L for cell B, and LOD 133.1680 ng/100 µL and 532.6720 ng/100 µL for cell A and cell B respectively with correlation coefficient (r) 0.9910 for cell A and 0.9901 for cell B, RSD% was lower than 1%, (n=8) for the determination of ate

... Show More
View Publication Preview PDF
Scopus (3)
Crossref (4)
Scopus Clarivate Crossref
Publication Date
Fri Nov 24 2023
Journal Name
Iraqi Journal Of Science
New Turbidimetric-Continuous Flow Injection Analysis Method for The Determination of Chlorpromazine HCl in Pharmaceutical Preparation Using Linear Array Ayah 5SX1-T-1D-CFI Analyser
...Show More Authors

A new turbidimetric-flow injection method is described for the determination of chlorpromazine HCl in pure and pharmaceutical preparation. The method is characterized by simplicity, sensitivity and fast, it is based on formation of ion pair compound between chlorpromazine HCl and Potassium hexacyanoferrate(III) in an acid medium for the formation of greenish yellow precipitate. This precipitate was determined using homemade Linear Array Ayah 5SX1-T-1D continuous flow injection analyser. Optimum concentrations of chemical reactants, physical instrumental conditions have been investigated. The linear dynamic range of chlorpromazine HCl was 3-30 mmol.L-1 while correlation coefficient (r) was 0.9929 and percentage linearity (%r2) C.O.D was 9

... Show More
View Publication Preview PDF
Publication Date
Mon May 15 2023
Journal Name
Iraqi Journal Of Science
Determination of hydrogen peroxide in some local pharmaceutical disinfectants by continuous flow injection analysis via turbidimetric (T180o) and scattered light effect at two opposite positions (2N90o ) using Ayah 4SW-3D-T180o -2N90o -Solar - CFI Analyse
...Show More Authors

An indirectly method is used to determine hydrogen peroxide. The method based on oxidation of chromium (III) ion by hydrogen peroxide in basic medium to form chromate ion which react with barium (II) ion to produce a yellow precipitate (BaCrO4). Under the optimum established conditions, the linear range of 0.50-25.00 mmol L-1 along with correlation coefficient (r) of 0.9992, Limit of detection (LOD) 0.68 μg / 100 μL, precision expressed as relative standard deviation for six replication measurements at 5.0 mmol.L-1 H2O2 of less than 2% were obtained for hydrogen peroxide. The developed method was successfully applied for the estimation of H2O2 in three pharmaceuticals preparation of different companies using continuous flow injection o

... Show More
View Publication Preview PDF
Publication Date
Sat Jan 01 2022
Journal Name
Chemical Methodologies
Determination of the Quantity of Losartan Active Ingredient in the Medication Formulations via Turbidimetric-Flow Injection Technique
...Show More Authors

Scopus (14)
Scopus
Publication Date
Fri Jul 01 2022
Journal Name
Iraqi Journal Of Science
New Approach for Determination of Ciprofloxacin by Quenched Fluorescence of Analytically Interested Species Using Continuous Flow Injection Laser Diode Fluorimeter Analyser
...Show More Authors

The objective of this study was to establish an accurate, precise, sensitive, simple, fast and reliable method for the determination of ciprofloxacin hydrochloride in pure or in pharmaceutical dosage forms using Homemade instrument fluorimeter continuous flow injection analyser with solid state laser (405 nm) as a source. The method is based upon the fluorescence fluorescein sodium salt and its quenching by ciprofloxacin hydrochloride in aqueous medium. The solutions of standard and the sample were prepared in distilled water. The calibration graph was linear in the concentration range of using (10 - 100) mMol.L-1 ciprofloxacin hydrochloride (r= 0.9891) with relative standard deviation (RSD%) for 3 mMol.L-1 ciprofloxacin HCl solution is

... Show More
View Publication Preview PDF
Publication Date
Tue Jan 01 2019
Journal Name
Indian Journal Of Public Health Research & Development
Spectrophotometric Determination of Nifedipine in Pharmaceutical Tablets Using Batch and Flow Injection Method Via Diazotization Coupling Reaction
...Show More Authors

View Publication
Scopus (2)
Crossref (1)
Scopus Crossref
Publication Date
Sun Oct 22 2023
Journal Name
Iraqi Journal Of Science
New Approach for the On-Line Spectrophotometric Determination of Vanadium (V) In Different River Water via the Use of a Homemade Ayah 6SX1-T-2D Solar-Continuous Flow Injection Analyser
...Show More Authors

A new simple sensitive and selective spectrophotometric method has been developed for the analysis of vanadium(V) in three randomly chosen samples from river water at different locations by continuous flow injection analysis. The method based on the oxidation of pyrogallol by vanadium(V) in acidic solution to form color species and the same species was determined using homemade Ayah 6SX1-T-2D solar cell analyser . Chemical and physical parameters were investigated using the high intensity of snow white light emitted diode as a source. The linear dynamic range for the instrument response versus vanadium(V) concentration was 1-200 mg.L-1 with correlation coefficient r = 0.9920. The limit of detection (S/N=3) was 70 ng/ sample from the step

... Show More
View Publication Preview PDF