Accuracy Assessment of Stonex X-300 Laser Scanner Cameras
...Show More Authors

Assessment the actual accuracy of laboratory devices prior to first use is very important to know the capabilities of such devices and employ them in multiple domains. As the manual of the device provides information and values in laboratory conditions for the accuracy of these devices, thus the actual evaluation process is necessary.

In this paper, the accuracy of laser scanner (stonex X-300) cameras were evaluated, so that those cameras attached to the device and lead supporting role in it. This is particularly because the device manual did not contain sufficient information about those cameras.

To know the accuracy when using these cameras in close range photogrammetry, laser scanning (stonex X-300) device is used to obtain photos of a board including (23) ground control points on it, those observed from two stations and adjusted by using equations of the 3D adjusted triangulation networks by lengths and angles (hybrid routine). (10) GCPs and (13) checkpoints were used to compare the Root Mean Square Error (RMSE) of checkpoints that result from using laser scan cameras with (RMSE) of the same checkpoints that result from using digital photos (Nikon 5200D).

The result of (RMSE) comparison was ) in the X direction, ) in the Y direction and ) in the Z direction.

 

Crossref
View Publication Preview PDF
Quick Preview PDF
Publication Date
Wed Jun 11 2003
Journal Name
Iraqi Journal Of Laser
Microdrop Size Measurement in a Mixing Unit Using Laser Technique
...Show More Authors

One of the troublesome duties in chemical industrial units is determining the instantaneous drop size distribution, which is created between two immiscible liquids within such units. In this work a complete system for measuring instantaneous droplet size is constructed. It consists of laser detection system (1mW He-Ne laser), drop generation system (turbine mixer unit), and microphotography system. Two immiscible liquids, water and kerosene were mixed together with different low volume fractions (0.0025, 0.02) of kerosene (as a dispersed phase) in water (as a continuous phase). The experiments were carried out at different rotational speed (1180- 2090 r.p.m) of the turbine mixer. The Sauter mean diameter of the drops was determined by la

... Show More
View Publication Preview PDF
Publication Date
Sun Dec 04 2011
Journal Name
Baghdad Science Journal
Effect of Laser Irradiation on the Optical Properties of SnO2 films Deposited by Post Oxidation of metal Films
...Show More Authors

Tin oxide films (SnO2) of thickness (1 ?m) are prepared on glass substrate by post oxidation of metal films technique. Films were irradiated with Nd:YAG double frequency laser of wavelength (532 nm) pulses of three energies (100, 500, 1000) mJ. The optical absorption, transmission, reflectance, refractive index and optical conductivity of these films are investigated in the UV-Vis region (200-900) nm. It was found that the average transmittance of the films is around (80%) at wavelength (550 nm) and showed high transmission (? 90 %) in the visible and near infrared region. The absorption edge shifts towards higher energies, which is due to the Moss-Burstien effect and it lies at (4 eV). The optical band gap increased with increasing of ene

... Show More
View Publication Preview PDF
Crossref
Publication Date
Tue Feb 12 2019
Journal Name
Iraqi Journal Of Physics
A Study of the generation of laser soliton from spontaneous emission by ring cavity with carbon nanotube
...Show More Authors

We demonstrate a behavior of laser pulse grows through fiber laser inside and output cavity with a soliton fiber laser based on the multi-wall carbon nanotube saturable absorber (SA), we investigate the effects of a saturable absorber parameter on the mode-locking of a realistic Erbium fiber ring laser. Generalized nonlinear Schrodinger equation including the nonlinear effects as gain dispersion, second anomalous group velocity dispersion (GVD), self phase modulation (SPM), and two photon absorption used to describe pulse evolution. An analytical method has been used to understand and to quantify the role of the SA parameter on the propagation dynamics of pulse laser. We compute the chirp, power, width and phase of the soliton for range

... Show More
View Publication Preview PDF
Crossref
Publication Date
Tue Nov 30 2021
Journal Name
Iraqi Journal Of Science
Synthesis and Characteristics of Ag, Cu/Au Core/Shell Nanoparticles Produced by Pulse Laser Ablation
...Show More Authors

Colloidal dispersions of mono Au, Ag , Cu and bimetallic Ag/Au and Cu/Au
core/shell nanoparticles are synthesized by pulsed laser ablation of metals targets
immersed in 5 ml distilled water (DW). Surface Plasmon resonance (SPR) and
particle sizes are characterized by UV-VIS and HRTEM, the X-ray diffraction
shows the structure of core/shell. The Surface Plasmon resonance of the produced
nanoparticles solutions for silver nanoparticles about 402 nm and copper
nanoparticles about 636 nm. While for the core-shell observed two peaks of SPR,
Ag/Au core/shell at (406-516) nm, and Cu/Au core/shell observed one peak at
565nm, because the region of gold and copper close together. The shape and
particle size have been con

... Show More
View Publication Preview PDF
Publication Date
Thu Apr 18 2019
Journal Name
Iraqi Journal Of Science
Spectral and Linear Optical Characterization of Rhodamine B and Fluorescein Sodium Organic Laser Dyes Mixture Solutions
...Show More Authors

Spectral and linear optical properties for a mixture of Rhodamine B (RB) and Fluorescein Sodium (Na Fl) organic laser dyes were determined at different concentrations 10-3, 10-4 M in ethanol solvent at room temperature. The intensity of absorption range is towards longer wavelengths (red shift). The quantum efficiency diminished while the radiative and fluorescence life time increased when increment concentration, organic laser dyes have a spectrum within the range 540-500 nm. Results demonstrate that a mixture of laser dyes are effective optical materials when contrasted with individual laser dyes. It can be utilized as resonator in cavity lasers.

View Publication Preview PDF
Publication Date
Wed Feb 08 2023
Journal Name
Iraqi Journal Of Science
Laser-Induced Modification of Ag and Cu Metal Nanoparticles Formed by Exploding Wire Technique in Liquid
...Show More Authors

This research aims to find a method to synthesize nanoparticles of important metals in the fields of medicine and electronics, with high purity small in size and narrow size distribution. And it characterized by simplicity, efficiency and high productivity. To achieve this aim the effects of laser irradiation on silver and copper colloids prepared by exploding wire technique in double distilled and deionized water (DDDW) have been studied. The laser irradiation was performed using laser radiation fluence about 4 J/cm2 at 532 nm wavelength. Additional irradiation of colloids resulted in the changes of particles morphology, which were monitored by absorption spectroscopy and transmission electron microscopy methods. It was found that both

... Show More
View Publication Preview PDF
Publication Date
Sun Oct 01 2023
Journal Name
Baghdad Science Journal
The Influence of Ablation Speed on the Synthesis of Carbon Nanostructures Via Pulsed Laser Ablation of Asphalt in Ethanol
...Show More Authors

Pulsed liquid laser ablation is considered a green method for the synthesis of nanostructures because there are no byproducts formed after the ablation. In this paper, a fiber laser of wavelength 1.064 µm, peak power of 1 mJ, pulse duration of 120 ns, and repetition rate of 20 kHz, was used to produce carbon nanostructures including carbon nanospheres and carbon nanorods from the ablation of asphalt in ethanol at ablation speeds of (100, 75, 50, 10 mm/s).  The morphology, composition and optical properties of the synthesized samples were studied experimentally using FESEM, HRTEM, EDS, and UV-vis spectrophotometer. Results showed that the band gap energy decreased with decreasing the ablation speed (increasing the ablation time), the mi

... Show More
View Publication Preview PDF
Scopus (1)
Crossref (1)
Scopus Crossref
Publication Date
Fri Jun 16 2023
Journal Name
Iraqi Journal Of Pharmaceutical Sciences ( P-issn 1683 - 3597 E-issn 2521 - 3512)
Role of Fasting Mimicking Diet in Farnesoid x Receptor for Suppressing Epithelial-to-Mesenchymal Transition, Cell Cycle Progression, and Viability of Prostate Cancer Cells
...Show More Authors

The systemic and resistant nature of metastatic castration-resistant prostate cancers (mCRPC) renders it largely incurable even after intensive multimodal therapy. Proliferation, survival, and epithelial-mesenchymal transition (EMT) are three fundamental events that are deeply linked to carcinogenesis.  Hence, it is necessary to find a new combination of several therapies, targeting those vital mechanisms without causing side effects. Significant research works have shown differential low expression of the metabolic Farnesoid X receptor (FXR) in primary and metastatic prostate cancer suggesting their importance in prostate pathogenesis. Obticholic acid (INT 747), a potent FXR agonist is widely used in primary biliary chola

... Show More
View Publication Preview PDF
Scopus (2)
Crossref (1)
Scopus Crossref
Publication Date
Sat Jan 01 2022
Journal Name
3rd International Scientific Conference Of Alkafeel University (iscku 2021)
Analysis of X-ray diffraction lines profile of Tio2 nanoparticles to determine the energy per unit volume and stress by using Halder-Wagner method
...Show More Authors

In this study, the Halder-Wagner method was used for an analysisX-ray lines of Tio2 nanoparticles. Where the software was used to calculate the FWHM and integral breath (β) to calculate the area under the curve for each of the lines of diffraction. After that, the general equation of the halder- Wagner method is applied to calculate the volume (D), strain (ε), stress (σ), and energy per unit(u). Volume (β). Where the value of the crystal volume was equal to (0.16149870 nm) and the strain was equal to (1.044126), stress (181.678 N / m2), and energy per unit volume (94.8474 J m-3).The results obtained from these methods were then compared with those obtained from each of the new paradigm of the HalderWagner method, the Shearer developm

... Show More
View Publication
Scopus Crossref
Publication Date
Sun Mar 06 2011
Journal Name
Baghdad Science Journal
Study of opacity broadening in spectral lines for helium like ions in aluminum plasma which produced by laser
...Show More Authors

A theoretical analysis studied was performed to study the opacity broadening of spectral lines emitted from aluminum plasma produced by Nd-YLF laser. The plasma density was in the range 1028-1026 )) m-3 with length of plasma about ?300) m) , the opacity was studied as function of plasma density & principle quantum number. The results show that the opacity broadening increases as plasma density increases & decreases with the spacing between energy levels of emission spectral line.

View Publication Preview PDF
Crossref