Chromatographic and spectrophotometric methods for the estimation of mebendazole in
pharmaceutical products were developed. The flow injection method was based on the oxidation of
mebendazole by a known excess of sodium hypochlorite at pH=9.5. The excess sodium hypochlorite is then
reacted with chloranilic acid (CAA) to bleach out its color. The absorbance of the excess CAA was recorded
at 530 nm. The method is fast, simple, selective, and sensitive. The chromatographic method was carried out
on a Varian C18 column. The mobile phase was a mixture of acetonitrile (ACN), methanol (MeOH), water
and triethylamine (TEA), (56% ACN, 20% MeOH, 23.5% H2O, 0.5% TEA, v/v), adjusted to pH = 3.0 with
1.0 M hydrochloric acid. Naphazoline nitrate was used as an internal standard. The absorption of mebendazole
was measured using a variable wavelength UV detector at 290 nm. Linearity was obtained in the concentration
range of 1-60 and 0.10-3.0 mg/L for the HPLC and FIA, respectively. The methods were applied successfully
for the assay of mebendazole in pharmaceutical products and no interferences were observed from the common
excipients usually used. The proposed methods were validated for their accuracy and precision.
A reliable differential pulse polarographic (DPP) method has been developed and applied for the determination of ibuprofen IBU in dosage form with dropping mercury electrode (DME) versus Ag/AgCl. The best peak was found at cathodic peak of -1.18 V in phosphate buffer at pH=4 and 0.025M of KNO3 as supporting electrolyte. In order to obtaine the highest sensitivity, instrumental and experimental parameters were examined including the type and concentration of supporting electrolyte, pH of buffer solution, pulse amplitude and voltage step time. Diffusion current showed a direct linear relationship to ibuprofen concentration in the range of (5 – 30) μg. mL-1 (2.43× 10-5
... Show MoreAmiodarone hydrochloride (AH) has been determined spectrophotometrically Using methyl orange (MO). In our previous researches MO was used for determination of Mexiletine Hydrochloride [1]. The method based on complexation between MO and AH. After shaking and diluting the complex solution with D.W, the pH was adjusted with NaOH and HCl to pH 3. The colored complex formed between AH and the reagent were transferred into separating funnels and extracted using 5.5ml CH2Cl2 and were shaken for (5 minutes). The extracted organic layer was used for preparation of the calibration curves for spectrophotometric measurements of AH at 434nm. The blanks were carried out in exactly the same way throughout the whole procedure.&n
... Show More A simple, accurate and precise spectrophotometric method has been proposed for the determination of Mefenamic acid(MA) in dosage forms. Proposed method based on the reaction of cited drug with 1,2-Naphthoquinone-4-Sulfonic sodium (NQS). The optimum experimental condition have been studied. Beer's Law is obeyed in the concentration range 0.5-10.0 µg/mL at 450nm with detection limit of 0.189µg/mL. Effect of pH, reaction time, and volume of NQS on the determination of Mefenamic acid, have been examined. The proposed method has been successfully applied for the determination of Mefenamic acid in pharmaceutical preparations.
A rapid and sensitive method for the determination of amoxicillin trihydrate (AMXT) based on the diazo-coupling reaction was studied. Sulphanilic acid diazotizes with nitrite ion in acidic medium to produce a water soluble, colorless diazonium ion, which subsequently coupled with AMXT to form a colored azo dye in the alkaline medium, having maximum absorption at 455 nm. The calibration graph showed that Beer's law is obeyed over the concentration range of 0.3 – 30.0 μg/mL of AMXT, with the detection limit of 0.15 μg/mL and molar absorptivity was 2.3 × 104 L/mol.cm. The accuracy and the precision were acceptable depending upon the values of error percentage and relative standard deviation. The influence of common interfer
... Show MoreFour rapid, accurate and very simple derivative spectrophotometric techniques were developed for the quantitative determination of binary mixtures of estradiol (E2) and progesterone (PRG) formulated as a capsule. Method I is the first derivative zero-crossing technique, derivative amplitudes were detected at the zero-crossing wavelength of 239.27 and 292.51 nm for the quantification of estradiol and 249.19 nm for Progesterone. Method II is ratio subtraction, progesterone was determined at λmax 240 nm after subtraction of interference exerted by estradiol. Method III is modified amplitude subtraction, which was established using derivative spectroscopy and mathematical manipulations. Method IIII is the absorbance ratio technique, absorba
... Show MoreNitroso-R-salt is proposed as a sensitive spectrophotometric reagent for the determination of paracetamol in aqueous solution. The method is based on the reaction of paracetamol with iron(III) and subsequent reaction with nitroso-R-salt to yield a green colored complex with maximum absorption at 720 nm. Optimization of the experimental conditions was described. The calibration graph was linear in the concentration range of 0.1 – 2.0 ?g mL-1 paracetamol with a molar absorptivity of 6.9 × 104 L mol-1 cm-1. The method was successfully applied to the determination of paracetamol in pharmaceutical preparations without any interference from common excipients. The method has been statistically evaluated with British Pharmacopoeia method a
... Show MoreVerapamil Hydrochloride (VH) has been determined spectrophotometrically by using Methyl Orange (MO). In our previous researches MO was used for determination of Mexiletine Hydrochloride [1]. The method was based on complexation between (MO and VH). After shaking and diluting the complex solution with D.W, the pH was adjusted with NaOH and HCl to pH 4. The colored complex formed between VH and the reagents were transferred into separating funnels and extracted using 4.5 ml CH2Cl2 and were shaken for (4 minutes). The extracted organic layer was used for the preparation of the calibration curves for spectrophotometric measurements of VH at 437nm. The blanks were carried out in exactly the same way throughout the whol
... Show MoreA simple, cheap, fast, accurate, Safety and sensitive spectrophotometric method for the determination of sulfamethaxazole (SFMx), in pure form and pharmaceutical dosage forms. has been described The Method is based on the diazotization of the drug by sodium nitrite in acidic medium at 5Cº followed by coupling with salbutamol sulphate (SBS) drug to form orange color the product was stabilized and measured at 452 nm Beer’s law is obeyed in the concentration range of 2.5-87.5 ?g ml-1 with molar absorptivity of 2.5x104 L mole-1 cm-1. All variables including the reagent concentration, reaction time, color stability period, and sulfamethaxazole /salbutamol ratio were studied in order to optimize the reaction conditions. No interferences were
... Show MoreThe first flow injection spectrophotometric method is characterized by its speed and sensitivity which have been developed for the determination of promethazine-HCl in pure and pharmaceutical preparation. It is based on the in situ detection of colored cationic radicals formed via oxidation of the drug with sodium persulphate to pinkish-red species and the same species was determined by using homemade Ayah 3SX3-3D solar flow injection photometer. Optimum conditions were obtained by using the high intensive green light emitted diode as a source. Linear dynamic range for the absorbance versus promethazine-HCl concentration was 0-7 mmol.L-1, with the correlation coefficient (r) was 0.9904 while the percentage linearity (r2%) was 98.09%. the L.
... Show MoreA simple and accurate method to determinate furosemide (FUR) based on converting the secondary amine to primary amine with acidic hydrolysis then azotization by nitrous acid and coupled with resorcinol as a coupling agent in aqueous medium at pH 13. The optical characteristic like beers law limit found to be (0.25-2.5) μg.ml−1, detection and quantification limits (0.0196) (0.0654) μg.ml−1respectivly and Sandel sensitivity was 0.006738 μg.cm−2. The least-square method was used to evaluate the regression equation and the correlation coefficient. The resulted azo dye has a maximum absorbance at 430 nm with light oran