Chromatographic and spectrophotometric methods for the estimation of mebendazole in
pharmaceutical products were developed. The flow injection method was based on the oxidation of
mebendazole by a known excess of sodium hypochlorite at pH=9.5. The excess sodium hypochlorite is then
reacted with chloranilic acid (CAA) to bleach out its color. The absorbance of the excess CAA was recorded
at 530 nm. The method is fast, simple, selective, and sensitive. The chromatographic method was carried out
on a Varian C18 column. The mobile phase was a mixture of acetonitrile (ACN), methanol (MeOH), water
and triethylamine (TEA), (56% ACN, 20% MeOH, 23.5% H2O, 0.5% TEA, v/v), adjusted to pH = 3.0 with
1.0 M hydrochloric acid. Naphazoline nitrate was used as an internal standard. The absorption of mebendazole
was measured using a variable wavelength UV detector at 290 nm. Linearity was obtained in the concentration
range of 1-60 and 0.10-3.0 mg/L for the HPLC and FIA, respectively. The methods were applied successfully
for the assay of mebendazole in pharmaceutical products and no interferences were observed from the common
excipients usually used. The proposed methods were validated for their accuracy and precision.
Because of their Physico‐chemical characteristics and its composition, the development of new specific analytical methodologies to determine some highly polar pesticides are required. The reported methods demand long analysis time, expensive instruments and prior extraction of pesticide for detection. The current work presents a new flow injection analysis method combined with indirect photometric detection for the determination of Fosetyl‐Aluminum (Fosetyl‐Al) in commercial formulations, with rapid and highly accurate determination involving only construction of manifold system combined with photometric detector without need some of the pre‐treatments to the sample before the analysis such a
A simple, accurate and sensitive spectrophotometric method for the determination of Procaine penicillin (PP) is described. The method is based on charge-transfer reaction of PP with metol (N-methyl-p-hydroxy aniline) in the presence of ferric sulphate to form a purple-water soluble complex ,which is stable and has a maximum absorption at 510 nm .A graph of absorbance versus concentration shows that Beer’s low is obeyed over the concentration range of 3-80 µg /ml of PP (i.e.,3-80 ppm) with a molar absorbativity of 4.945 ×103 L.mol-1.cm-1 ,Sandell sensitivity of 0.1190 µg cm-2 ,a relative error of (-1.57)-2.79 % and a standard deviation of less than 0.59 depending on the concentration of PP.The optimum conditions for full co
... Show MoreSimple and sensitive batch and Flow-injection spectrophotometric methods for the determination of Procaine HCl in pure form and in injections were proposed. These methods were based on a diazotization reaction of procaine HCl with sodium nitrite and hydrochloric acid to form diazonium salt, which is coupled with chromatropic acid in alkaline medium to form an intense pink water-soluble dye that is stable and has a maximum absorption at 508 nm. A graphs of absorbance versus concentration show that Beer’s law is obeyed over the concentration range of 1-40 and 5-400 µg.ml-1 of Procaine HCl, with detection limits of 0.874 and 3.75 µg.ml-1 of Procaine HCl for batch and FIA methods respectively. The FIA average sample throughput was 70 h-1. A
... Show MoreSimple and sensitive kinetic methods are developed for the determination of Paracetamol in pure form and in pharmaceutical preparations. The methods are based on direct reaction (oxidative-coupling reaction) of Paracetamol with o-cresol in the presence of sodium periodate in alkaline medium, to form an intense blue-water-soluble dye that is stable at room temperature, and was followed spectrophotometriclly at λmax= 612 nm. The reaction was studied kinetically by Initial rate and fixed time (at 25 minutes) methods, and the optimization of conditions were fixed. The calibration graphs for drug determination were linear in the concentration ranges (1-7 μg.ml-1) for the initial rate and (1-10 μg.ml-1) for the fixed time methods at 25 min.
... Show MoreA new spectrophotometric flow injection method has been establish for the determintaions of some catecholaminedrugs
The aim of this work was to develop and validate a rapid and low cost method for estimation of ibuprofen in pharmaceutical suspensions using Reverse-Phase High Performance Liquid Chromatography. The proposed method was conducted and validated according to International Conference on Harmonization (ICH) requirements. The chromatographic parameters were as follows: column of octyldecylsilyl C18 with dimensions (150 × 4.6) mm, mobile phase composed of acetonitrile with phosphoric acid with a ratio of 50 to 50 each using isocratic mode, flow rate of 1.5 mL/min and injection volume of 5 μL. The detection was carried out using UV detector at 220 nm. The method was validated and showed short retention time for ibuprofen peak at 7.651 min, wit
... Show MoreGenistein (GEN) is The major isoflavone found in soybeans, has a number of cardiovascular health benefits, Postmenopausal syndrome and osteoporosis. A direct flow injection analysis method for estimation of (GEN) in pure and supplements formulation . This system is based on diazotization coupling reactions between procaine penciline (PR) and genistein in basic medium, they formed yellow dyes have maximum absorption at 416 nm. Calibration curve were constructed over different GEN concentrations, linearity for GEN was 10-100 µg.mL-1 and detection limits of 1.51 ?g/mL. In the FIA technique, all analytical factors were analyzed and optimized. The established method was successfully used to determine GEN in the formulations of its supplement
... Show MoreThis approach was developed to achieve an accurate, fast, economic and sensitivity to estimation of diphenhydramine Hydrochloride. The dye that produced via reaction between diphenhydramine HCl with thymol blue in acidic medium pH ≈ 4.0. The ion pair method include an optimization study to formed yellowcolored that extraction by liquid – liquid method. The product separated of complexes by using by chloroform solution measured spectrophotometry at 400 nm. The analysis data at optimum conditions showed that linearity concentration in a range of calibration curve 1.0 – 50 μg /mL, limit of detectionand limit of quantification 0.0786 and 0.2358 μg/mL respectively. The molar absorptivity and Sandell’s sensitivity were 1.8 × 10 -4 L/mo
... Show MoreA batch and flow injection (FI) spectrophotometric methods are described for the determination of barbituric acid in aqueous and urine samples. The method is based on the oxidative coupling reaction of barbituric acid with 4-aminoantipyrine and potassium iodate to form purple water soluble stable product at λ 510 nm. Good linearity for both methods was obtained ranging from 2 to 60 μg mL−1, 5–100 μg mL−1 for batch and FI techniques, respectively. The limit of detection (signal/noise = 3) of 0.45 μg mL−1 for batch method and 0.48 μg mL−1 for FI analysis was obtained. The proposed methods were applied successfully for the determination of barbituric acid in tap water, river water, and urine samples with good recoveries of 99.92
... Show More