Chromatographic and spectrophotometric methods for the estimation of mebendazole in
pharmaceutical products were developed. The flow injection method was based on the oxidation of
mebendazole by a known excess of sodium hypochlorite at pH=9.5. The excess sodium hypochlorite is then
reacted with chloranilic acid (CAA) to bleach out its color. The absorbance of the excess CAA was recorded
at 530 nm. The method is fast, simple, selective, and sensitive. The chromatographic method was carried out
on a Varian C18 column. The mobile phase was a mixture of acetonitrile (ACN), methanol (MeOH), water
and triethylamine (TEA), (56% ACN, 20% MeOH, 23.5% H2O, 0.5% TEA, v/v), adjusted to pH = 3.0 with
1.0 M hydrochloric acid. Naphazoline nitrate was used as an internal standard. The absorption of mebendazole
was measured using a variable wavelength UV detector at 290 nm. Linearity was obtained in the concentration
range of 1-60 and 0.10-3.0 mg/L for the HPLC and FIA, respectively. The methods were applied successfully
for the assay of mebendazole in pharmaceutical products and no interferences were observed from the common
excipients usually used. The proposed methods were validated for their accuracy and precision.
Because of their Physico‐chemical characteristics and its composition, the development of new specific analytical methodologies to determine some highly polar pesticides are required. The reported methods demand long analysis time, expensive instruments and prior extraction of pesticide for detection. The current work presents a new flow injection analysis method combined with indirect photometric detection for the determination of Fosetyl‐Aluminum (Fosetyl‐Al) in commercial formulations, with rapid and highly accurate determination involving only construction of manifold system combined with photometric detector without need some of the pre‐treatments to the sample before the analysis such a
In this research, salbutamol sulphate (SAS) has been determined by a simple, rapid and sensitive spectrophotometric method. Salbutamol sulphate in this method is based on the coupling of SAS with diazotized ρ- bromoaniline reagent in alkaline medium of Triton X-100 (Tx) to form an orange azo dye which is stable and water-soluble. The azo dye is exhibiting maximum absorption at 441 nm. A 10 - 800 µg of SAS is obeyed of Beer's law in a final volume of 20 ml, i.e., 0.5- 40 ppm with ε, the molar absorptivity of 48558 L.mol-1.cm-1 and Sandell's sensitivity index of 0.01188 µg.cm-2. This new method does not need solvent extraction or temperature control which is well applied to determine SAS in d
... Show MoreSimple and sensitive batch and flow injection methods (normal and reverse flow injection analysis (nFIA and rFIA)) for spectrophotometric determination of vancomycin hydrochloride (VHC) in pharmaceutical preparations were proposed and optimized. Both methods are based on the oxidative - coupling reaction between vancomycin hydrochloride and 2,4-Dinitrophenylhydrazine (DNPH) in the presence of sodium periodate in alkaline medium to form a yellow water-soluble product that is stable and has a maximum absorption at 461 nm. Beer’s law was obeyed over the range of 1- 40, 0.5-120 and 0.5-150 μg.mL-1; the limits of detection were 0.537, 0.0823 and 0.233 μg.mL-1 for batch, normal and reverse flow injection methods respectively. The sampling
... Show MoreA simple indirect spectrophotometric method for determination of mebendazol in pure and pharmaceutical formulation was presented in this study. UV-Visible spectrophotometry using the optimal conditions was developed for determination of mebendazole in pure drug and different preparation samples. The method is based on the oxidation of drug by nbromosuccinimide with hydrochloric acid and the left amount of oxidizing agent was determined by the reaction with tartarazine and the absorbance was measured at 428 nm. Calibration curves were linear in the range of 5 to 30 µg.mL-1 with molar absorptivity 8437.2 L.mol-1 .cm-1 . The limits of detection and quantification were determined and found to be 0.7770 µg.mL-1 and 2.3400 µg.mL-1 respec
... Show MoreA simple, accurate and sensitive spectrophotometric method for the determination of Procaine penicillin (PP) is described. The method is based on charge-transfer reaction of PP with metol (N-methyl-p-hydroxy aniline) in the presence of ferric sulphate to form a purple-water soluble complex ,which is stable and has a maximum absorption at 510 nm .A graph of absorbance versus concentration shows that Beer’s low is obeyed over the concentration range of 3-80 µg /ml of PP (i.e.,3-80 ppm) with a molar absorbativity of 4.945 ×103 L.mol-1.cm-1 ,Sandell sensitivity of 0.1190 µg cm-2 ,a relative error of (-1.57)-2.79 % and a standard deviation of less than 0.59 depending on the concentration of PP.The optimum conditions for full co
... Show MoreSimple and sensitive batch and Flow-injection spectrophotometric methods for the determination of Procaine HCl in pure form and in injections were proposed. These methods were based on a diazotization reaction of procaine HCl with sodium nitrite and hydrochloric acid to form diazonium salt, which is coupled with chromatropic acid in alkaline medium to form an intense pink water-soluble dye that is stable and has a maximum absorption at 508 nm. A graphs of absorbance versus concentration show that Beer’s law is obeyed over the concentration range of 1-40 and 5-400 µg.ml-1 of Procaine HCl, with detection limits of 0.874 and 3.75 µg.ml-1 of Procaine HCl for batch and FIA methods respectively. The FIA average sample throughput was 70 h-1. A
... Show MoreThe aim of this work was to develop and validate a rapid and low cost method for estimation of ibuprofen in pharmaceutical suspensions using Reverse-Phase High Performance Liquid Chromatography. The proposed method was conducted and validated according to International Conference on Harmonization (ICH) requirements. The chromatographic parameters were as follows: column of octyldecylsilyl C18 with dimensions (150 × 4.6) mm, mobile phase composed of acetonitrile with phosphoric acid with a ratio of 50 to 50 each using isocratic mode, flow rate of 1.5 mL/min and injection volume of 5 μL. The detection was carried out using UV detector at 220 nm. The method was validated and showed short retention time for ibuprofen peak at 7.651 min, wit
... Show MoreDetermining the actual amounts of active ingredients in various pharmaceutical commercial forms is still receiving a lot of attention. Two flow injection analysis (FIA) methods were suggested for determination of mesalazine (MES) in pharmaceutical forms. Normal and reverse FIA systems (nFIA and rFIA) combined with UV-Vis spectrophotometric technique were used for the analysis. The methods involved using two mods of FIA systems for measuring a colored product result from coupling of MES with 2,2'-dihydroxybiphenyl after oxidized with sodium periodate in alkaline medium. The absorbance of the red colored dye was measured at maximum wavelength of 500 nm. The calibration graphs for MES were linear in the ranges 2.5-200 and 0.5-60 µg/mL with