Simple, sensitive and accurate two methods were described for the determination of terazosin. The spectrophotometric method (A) is based on measuring the spectral absorption of the ion-pair complex formed between terazosin with eosin Y in the acetate buffer medium pH 3 at 545 nm. Method (B) is based on the quantitative quenching effect of terazosin on the native fluorescence of Eosin Y at the pH 3. The quenching of the fluorescence of Eosin Y was measured at 556 nm after excitation at 345 nm. The two methods obeyed Beer’s law over the concentration ranges of 0.1-8 and 0.05-7 µg/mL for method A and B respectively. Both methods succeeded in the determination of terazosin in its tablets
The drug promethazine hydrochloride (PRZH) forms with rhodium (II) a colored chelate (?max = 472 nm) complex at (pH = 2.1) which is extractable with benzyl alcohol as organic solvent. Under the appropriate experimental conditions a calibration plot was set up from which some analytical parameter were derived and deduced by regression. Standard addition procedure was also adopted. It has been estimated that the concentration of the drug PRZH to be 24.89 mg per unit and 24.19 mg per unit for both calibrations. Under optimal conditions, the developed method has been achieved the following characteristics: LDR (30 – 150 µg ml-1 ) PRZH , RSD % ( 0.6 – 2.47 ) , sandell sensitivity( 0.0844 µg. cm -2 ) , LOD ( 1.66 µgml-1 ) , recovery
... Show MoreA new simple and sensitive spectrophotometric method for the determination of trace amount of Co(II) in the ethanol absolute solution have been developed. The method is based on the reaction of Co(II) with ethyl cyano(2-methyl carboxylate phenyl azo acetate) (ECA) in acid medium of hydrochloric acid (0.1 M) givining maximum absorbance at ((λmax = 656 nm). Beer's law is obeyed over the concentration range (5-60) (μg / ml) with molar absorptivity of (1.5263 × 103 L mol-1 cm-1) and correlation coefficient (0.9995). The precision (RSD% ˂ 1%). The stoichiometry of complex was confirmed by Job's method which indicated the ratio of metal to reagent is (2:1). The studied effect of interference elements Zn(II), Cu(II), Na(I), K(I), Ca(II) and Mg
... Show MoreIn this study, the stress-strength model R = P(Y < X < Z) is discussed as an important parts of reliability system by assuming that the random variables follow Invers Rayleigh Distribution. Some traditional estimation methods are used to estimate the parameters namely; Maximum Likelihood, Moment method, and Uniformly Minimum Variance Unbiased estimator and Shrinkage estimator using three types of shrinkage weight factors. As well as, Monte Carlo simulation are used to compare the estimation methods based on mean squared error criteria.
Verapamil Hydrochloride (VH) has been determined spectrophotometrically by using Methyl Orange (MO). In our previous researches MO was used for determination of Mexiletine Hydrochloride [1]. The method was based on complexation between (MO and VH). After shaking and diluting the complex solution with D.W, the pH was adjusted with NaOH and HCl to pH 4. The colored complex formed between VH and the reagents were transferred into separating funnels and extracted using 4.5 ml CH2Cl2 and were shaken for (4 minutes). The extracted organic layer was used for the preparation of the calibration curves for spectrophotometric measurements of VH at 437nm. The blanks were carried out in exactly the same way throughout the whol
... Show MoreA simple, sensitive, accurate and low cost effective spectrophotometric method has been developed for the determination of Tetracycline and Doxycycline in pure and pharmaceutical formulations. The method is based on the reaction of methyldopa with 4-aminoantipyren (4-AAP) in presence of potassium ferriecyanide (PFC) in an alkaline medium. Two optimization methods were applied to determine the optimum conditions of oxidizing coupling reaction variables; univariate and design of experiment (DOE) method. The conditions effecting the reaction; pH, buffer Volume, reagent concentration, oxidant concentration, type of buffer, order of addition, time of reaction and stability were optimized . Under univariate and design
... Show MoreA simple, fast and sensitive spectrophotometric method has been applied for the determination of tetracycline hydrochloride in its pure form and in pharmaceutical preparations. The method based on coupling reaction of the antibiotic with diazotized anthranilic acid to form a stable yellow azo dye which shows a maximum absorption at 419 nm. Uni- and multivariate approaches were followed in optimizing the experimental conditions. Under optimum experimental conditions obtained via multivariate (Central Composite Design), the linearity of the constructed calibration curve was in the range of 0.560 μg.mL-1 with molar absorptivity of 14619 L.mol-1.cm-1 and the value of detection limit was 0.2813μg.mL-1. The capability of the metho
... Show MoreA procedure for the mutual derivatization and determination of thymol and Dapsone was developed and validated in this study. Dapsone was used as the derivatizing agent for the determination of thymol, and thymol was used as the derivatizing agent for the determination of Dapsone. An optimization study was performed for the derivatization reaction; i.e., the diazonium coupling reaction. Linear regression calibration plots for thymol and Dapsone in the direct reaction were constructed at 460 nm, within the concentration range of 0.3-7 μg ml-1 for thymol and 0.3-4 μg ml-1 for Dapsone, with limits of detection 0.086 and 0.053 μg ml-1, respectively. Corresponding plots for the cloud point extraction of thymol and Dapsone were constructed
... Show MoreA simple, cheap, fast, accurate, Safety and sensitive spectrophotometric method for the determination of sulfamethaxazole (SFMx), in pure form and pharmaceutical dosage forms. has been described The Method is based on the diazotization of the drug by sodium nitrite in acidic medium at 5Cº followed by coupling with salbutamol sulphate (SBS) drug to form orange color the product was stabilized and measured at 452 nm Beer’s law is obeyed in the concentration range of 2.5-87.5 ?g ml-1 with molar absorptivity of 2.5x104 L mole-1 cm-1. All variables including the reagent concentration, reaction time, color stability period, and sulfamethaxazole /salbutamol ratio were studied in order to optimize the reaction conditions. No interferences were
... Show MoreSpectrophotometric methods were developed for the determination of rantidine-HCl in pharmaceutical tablets. These methods were based on the reaction of DDQ and p-chloranil with rantidine-HCl, resulting in the formation of an orange-red and purple colored products which are quantified spectrophotometrically at 460 and 540nm in DDQ and p-chloranil, respectively. A graph of absorbance versus concentration show that Beer’s law is obeyed in a concentration ranges of 20-160 and (30-120)ïg/ml with molar absorptivities of 2.631 x 103 and 1.052 x 103l .mol-1-cm-1 for DDQ and p-chloranil, respectively. The optimum conditions for color development are described and
... Show More