A direct, sensitive and efficient spectrophotometric method for the determination of nitrofurantoin
drug (NIT) in pure as well as in dosage form (capsules) was described. The suggested method was
based on reduction NIT drug using Zn/HCl and then coupling with 3-methyl-2-benzothiazolinone
hydrazone hydrochloride (MBTH) in the presence of ammonium ceric sulfate. Spectrophotometric
measurement was established by recording the absorbance of the green colored product at 610 nm.
Using the optimized reaction conditions, beer’s law was obeyed in the range of 0.5-30 μg/mL, with
good correlation coefficient of 0.9998 and limits of detection and quantitation of 0.163 and 0.544
μg/mL, respectively. The accuracy and precision of the proposed method represented by recovery and
relative standard deviation were satisfactory; about 99.33% and 1.16%, respectively. The proposed
method was applied for determination of NIT in its pharmaceutical forms and the results compared
successfully with those obtained by standard method (British pharmacopeia method).
Keywords: Nitrofurantoin , MBTH , Oxidative coupling reaction.
This researchpaper includes the incorporation of Alliin at various energy levels and angles
With Metformin using Gaussian 09 and Gaussian view 06. Two computers were used in this work. Samples were generated to draw, integrate, simulate and measure the value of the potential energy surface by means of which the lowest energy value was (-1227.408au). The best correlation compound was achieved between Alliin and Metformin through the low energy values where the best place for metformin to b
... Show MoreA simple, precise, rapid, and accurate reversed – phase high performance liquid chromatographic method has been developed for the determination of guaifenesin in pure from pharmaceutical formulations.andindustrial effluent. Chromatography was carried out on supelco L7 reversed- phase column (25cm × 4.6mm), 5 microns, using a mixture of methanol –acetonitrile-water: (80: 10 :10 v/v/v) as a mobile phase at a flow rate of 1.0 ml.min-1. Detection was performed at 254nm at ambient temperature. The retention time for guaifenesin was found 2.4 minutes. The calibration curve was linear (r= 0.9998) over a concentration range from 0.08 to 0.8mg/ml. Limit of detection (LOD) and limit of quantification ( LOQ) were found 6µg/ml and 18µg/ml res
... Show MoreThis study was undertaken to introduce a fast, accurate, selective, simple and environment-friendly colorimetric method to determine iron (II) concentration in different lipstick brands imported or manufactured locally in Baghdad, Iraq. The samples were collected from 500-Iraqi dinars stores to establish routine tests using the spectrophotometric method and compared with a new microfluidic paper-based analytical device (µPAD) platform as an alternative to cost-effective conventional instrumentation such as Atomic Absorption Spectroscopy (AAS). This method depends on the reaction between iron (II) with iron(II) selective chelator 1, 10-phenanthroline(phen) in the presence of reducing agent hydroxylamine (HOA) and sodium acetate (NaOAc) b
... Show MoreLiquid membrane electrodes for the determination iron(III) were constructed based on chloramphenicol sodium succinate and iron(III) CPSS-Fe(III) as ion pair complex, with four plasticizers Di-butyl phosphate (DBP); Di-butyl phthalate (DBPH); Di-octyl phthalate (DOP); Tri-butyl phosphate (TBP); in PVC matrix . These electrodes give Nernstian and sub-Nernstian slopes (19.79, 24.60, 16.01 and 13.82mV/decade) and linear ranges from (1x10-5-1x10-2 M, 1x10-5-1x10-2 M, 1x10-6-1x10-2 M and 1x10-5-1x10-2 M) respectively. The best electrode was based on DBP plasticizer which gave a slope 19.79 mV/decade, correlation coefficient 0.9999, detection limit of 9×10-6 M, lifetime 37 day displayed good stability and reproducibility and used to determine
... Show MoreTwo different approaches, univariate and multivariate (simplex method), have been used to obtain the optimum conditions for the quantitative Spectrophotometric determination of Eu3+ using Solochrome violet RS (3-Hydroxy-4-(2-hydroxy phenyl azo) naphthalene -1sulfonic acid) (SVRS) as a chromogenic reagent. The investigation shows that Eu3+ ion forms a wine-red complex with SVRS in alkaline buffer solution having a maximum absorbance at 464 nm against reagent blank. Calibration graphs obtained under univariate and simplex were found to be linear in the range of (0.30-8.0) µg/ml with detection limit 0.061µg/ml and molar absorptivity of 9877.66 L/mol.cm and (0.40-10.0)µg/ml with
... Show MoreThis work presents a simple method for determination of the neutron reflection coefficient (n) as a function of different neutron reflector materials.A laboratory neutron source (Am-Be) with activity of 16 ci is employed with a (BF3) neutron detector. Am-BeThree types of reflector materials are used as samples, the thickness of each sample is (5cm).It is found that( ?7) is: -For polyethlyene = 0.818
A specific, sensitive and simple method was used for the determination of: vitamin B9 (Folic acid) in pure and pharmaceutical formulations using continuous flow injection analysis. The method is based on formation of ion pair compound between folic acid and ammonium molybdate in an aqueous medium to obtain a gray precipitate complex, using homemade; Ayah-6SX1-ST-2D solar cell CFI Analyzer. Optimum parameters was studied to increase the sensitivity for developed method. The linear range for the calibration graph was 0.01-0.6 mMol.L-1 of vitamin B9 and LOD was 131.994 ng/sample with correlation coefficient ( r ) of 0.9810, RSD% was lower than 0.1%, (n=9) for the determination of vitamin B9 at concentration (0.07and 0.5) mMol.L-1 respectiv
... Show MoreThe aim of this work was to develop and validate a rapid and low cost method for estimation of ibuprofen in pharmaceutical suspensions using Reverse-Phase High Performance Liquid Chromatography. The proposed method was conducted and validated according to International Conference on Harmonization (ICH) requirements. The chromatographic parameters were as follows: column of octyldecylsilyl C18 with dimensions (150 × 4.6) mm, mobile phase composed of acetonitrile with phosphoric acid with a ratio of 50 to 50 each using isocratic mode, flow rate of 1.5 mL/min and injection volume of 5 μL. The detection was carried out using UV detector at 220 nm. The method was validated and showed short retention time for ibuprofen peak at 7.651 min, wit
... Show MoreA simple, rapid, accurate and sensitive spectrophotometric method is proposed for the detennination of chlorprQm<tZine -HCl in pwe form and in pharmaceutical formulation. This method is based on the formation. of ion assodation complexes of dmg with either thymol
blue or bromophenol blue in an acidic buffer at pH values 4.17 and
3.68, respectively.
The ion-pair complexes formed exhibit absorption maxima at 41 Onm for both thymol blue and bromophenol blue. These complexes· are quantitatively extracted &n
... Show MoreA simple analytical method was used in the present work for the simultaneous quantification of Ciprofloxacin and Isoniazid in pharmaceutical preparations. UV-Visible spectrophotometry has been applied to quantify these compounds in pure and mixture solutions using the first-order derivative method. The method depends on the first derivative spectrophotometry using zero-cross, peak to baseline, peak to peak and peak area measurements. Good linearity was shown in the concentration range of 2 to 24 µg∙mL-1 for Ciprofloxacin and 2 to 22 µg∙mL-1 for Isoniazid in the mixture, and the correlation coefficients were 0.9990 and 0.9989 respectively using peak area mode. The limits of detection (LOD) and limits of quantification (LOQ) were
... Show More