In this paper, turbidimetric and reversed-phase ultra-fast liquid chromatography (UFLC) methods were described for the quantitative determination of ephedrine hydrochloride in pharmaceutical injections form. The first method is based on measuring the turbidimetric values for the formed yellowish white precipitate in suspension status in order to determine the ephedrine hydrochloride concentration. The suspended substance is formed as a result of the reaction of ephedrine hydrochloride with phosphomolybdic acid which was used as a reagent. The physical and chemical characteristics of the complex were investigated. The calibration graphs of ephedrine were established by turbidity method. While the second method (UFLC) was conducted using the methanol-water (55+45, v/v) as the mobile phase with adjusted water pH 3.5. The ephedrine hydrochloride was detected and measured using UV detector at 260 nm. The linearity of ephedrine was obtained in the range of 0.09–0.39 mmol·l-1 . The detection limits (LOD) for the ephedrine hydrochloride were found to be 0.4 and 0.0044 mmol·l-1 by turbidity and UFLC, respectively. The developed methods were successfully applied for the quantitative determination of ephedrine hydrochloride in laboratory preparations (standard) and in commercial pharmaceutical injections. The two methods have given relative standard deviations (R.S.D.) in the range of 0.65–1.69 %, which indicates reasonable repeatability and high precision of both methods.
A simple, fast and sensitive spectrophotometric method has been applied for the determination of tetracycline hydrochloride in its pure form and in pharmaceutical preparations. The method based on coupling reaction of the antibiotic with diazotized anthranilic acid to form a stable yellow azo dye which shows a maximum absorption at 419 nm. Uni- and multivariate approaches were followed in optimizing the experimental conditions. Under optimum experimental conditions obtained via multivariate (Central Composite Design), the linearity of the constructed calibration curve was in the range of 0.560 μg.mL-1 with molar absorptivity of 14619 L.mol-1.cm-1 and the value of detection limit was 0.2813μg.mL-1. The capability of the metho
... Show Morea simple accurate and sensitive spectrophotometric method for the determination of promethazine HCI has been developed the method is based on the oxidative coupling reaction of promethazine
Amiodarone hydrochloride (AH) has been determined spectrophotometrically Using methyl orange (MO). In our previous researches MO was used for determination of Mexiletine Hydrochloride [1]. The method based on complexation between MO and AH. After shaking and diluting the complex solution with D.W, the pH was adjusted with NaOH and HCl to pH 3. The colored complex formed between AH and the reagent were transferred into separating funnels and extracted using 5.5ml CH2Cl2 and were shaken for (5 minutes). The extracted organic layer was used for preparation of the calibration curves for spectrophotometric measurements of AH at 434nm. The blanks were carried out in exactly the same way throughout the whole procedure.&n
... Show MoreDevelopment of a precise and delicate reaction has been acquired for the determination of vancomycin hydrochloride using batch and cloud point extraction (CPE) methods. The first method is based on the formation of azo dye as a result of diazotized dapsone coupled with vancomycin HCl (VAN) in a basic medium. The sensitivity of this reaction was enhanced by utilizing a nonionic surfactant (Triton X-114) and the cloud point extraction technique (second method). The azo dye formed was extracted into the surfactant-rich phase, dissolved in ethanol and detected at λmax 440 nm spectrophotometrically. The reaction was investigated using both batch and CPE methods (with and without extraction), and a simple comparison between the two
... Show MoreA simple, fast, and sensitive spectrophotometric method was suggested for the determination of Bromhexine Hydrochloride (BHH) in its pharmaceutical formulations. The method depends on the diazotization of BHH by sodium nitrite in acidic medium to produce the corresponding diazonium salt. The latter is coupled with phloroglucinol reagent in alkali medium to form a yellow water soluble azo-dye which has a maximum absorption at 405 nm with a molar absorptivity of 2.7×104 l.mol-1.cm-1 and Sandellʼs sensitivity of 0.01517 µg.cm-1. Beerʼs low is obeyed within a concentration range of 0.25-15 µg.mL-1 of BHH. The LOD and LOQ values of the proposed method were 0.087 µg.mL
Verapamil Hydrochloride (VH) has been determined spectrophotometrically by using Methyl Orange (MO). In our previous researches MO was used for determination of Mexiletine Hydrochloride [1]. The method was based on complexation between (MO and VH). After shaking and diluting the complex solution with D.W, the pH was adjusted with NaOH and HCl to pH 4. The colored complex formed between VH and the reagents were transferred into separating funnels and extracted using 4.5 ml CH2Cl2 and were shaken for (4 minutes). The extracted organic layer was used for the preparation of the calibration curves for spectrophotometric measurements of VH at 437nm. The blanks were carried out in exactly the same way throughout the whol
... Show MoreA simple, sensitive and accurate spectrophotometric method has been developed for the determination of salbutamol sulphate (SAB) and isoxsuprine hydrochloride (ISX) in pure and pharmaceutical dosage. The method involved oxidation of (SAB) and (ISX) with a known excess of N-bromosuccinamid in acidic medium, and subsequent occupation of unreacted oxidant in decolorization of Evans blue dye (EB). This, in the presence of SAB or ISX was rectilinear over the ranges 1.0-12.0, 1.0-11.0 µg/mL, with molar absorptivity 4.21×104 and 2.58×104 l.mol-1.cm-1 respectively. The developed method had been successfully applied for the determination of the studied drugs in their pharmaceutical dosage resulting i
... Show MoreIsocratic high performance liquid chromatography on reversed phase a (150x 4.6 mm I.D), 5 ?m ?-Bondapak RP-8 column (with acidic mobile phase allow the separation of doxcycycline hydrochloride with low detection limit of 0.2 µg/ml detected by UV set at 226 nm. The method was validated for Doxycycline between 0.156- to 5 µg/ml. The concentration of doxycycline was assessed in two single dose randomized crossover studies with intervals of one week between two period. In sera of 20 adults healthy male volunteers with average age of (42 + 10) year, body weight 48-85 kg, body height of (160-185cm) after a single dose of doxycycline hydrochoride 100 mg in form of capsules were orally administrated for both formulations. The blood sa
... Show MoreThe aim of this work was to develop and validate a rapid and low cost method for estimation of ibuprofen in pharmaceutical suspensions using Reverse-Phase High Performance Liquid Chromatography. The proposed method was conducted and validated according to International Conference on Harmonization (ICH) requirements. The chromatographic parameters were as follows: column of octyldecylsilyl C18 with dimensions (150 × 4.6) mm, mobile phase composed of acetonitrile with phosphoric acid with a ratio of 50 to 50 each using isocratic mode, flow rate of 1.5 mL/min and injection volume of 5 μL. The detection was carried out using UV detector at 220 nm. The method was validated and showed short retention time for ibuprofen peak at 7.651 min, wit
... Show MoreA new, effective, simple and inexpensive method was developed for determination the amounts of Ciprofloxacin HCl in solutions and in tablets by spectrophotometry and turbidity using sodium tungstate as reagent. Also, in this paper comparison between two techniques was conducted. The reaction between the Ciprofloxacin HCl and reagent in acidic media formed a yellowish white complex. Chemical and physical conditions have been investigated. The Linear range for spectroscopic and turbidimetric techniques were (0.05-1.25) and (0.05-3.0) mmol.L-1 with correlation coefficients (0.9956) and (0.9941) respectively. The limits of detection were 7×10-5 and 5×10-5 mol.L-1. The proposed method is good alternative for determination of Ciprofloxacin H
... Show More