A solid Phase Extraction (SPE) followed by HPLC-UV method is described for the simultaneous quantitative determination of nine priority pollutant phenols : Phenol, 2- and 4-Nitrophenol, 2,4-Dimethylphenol, 2-, 2,4-Di-, 2,4,6-Tri-, and Penta- chlorophenol, 4 Chloro-3-methylphenol. The phenols were separated using a C-18 column with UV detector at wave length of 280nm. The Flow of mobile phase was isocratic consisted of 50:50 Acetonitrile: phosphate buffer pH=7.1, column temperature 45 C°, Flow Rate 0.7 ml/min. Calibration curves were linear (R2 = 0.9961-0.9995). The RSDs (1.301-5.805)%, LOD(39.1- 412.4) µg/L, LOQ(118.5-1250.8) µg/L, the Robustness (1.55-4.89), Ruggedness (2.82-4.00), Repeatability (2.1-4.95), Recoveries% were (97.93 105.57)%.Condition of Extraction by (SPE) cartridges were optimized , the resin used is polystyrene-divinyl benzine, pH(2.2) and Elution solvent used is Tetrahydrofuran. The water samples to be analyzed were taken from six different locations. Three of them inside the station of petroleum refinery waste water treatment unit and the other three were at the Tigris River in Iraq around the station. The results were between (non detected, 1865.61) µg/L inside the station and (non detected, 374.66) µg/L at the river.