Simultaneous Determination of Trace Mefenamic Acid in Pharmaceutical Samples via Flow Injection Fluorometry
...Show More Authors

Mefenamic acid belongs to non-steroidal anti-inflammatory drugs that are used widely for the treatment of analgesia. Our aim from this study is to establish a new assay for the quantitative determination of mefenamic acid (MFA) in the pharmaceutical sample by two sensitive and rapid flow injection-fluorometric methods. A homemade fluorometer was used in fluorescence measurements, which using solid-state laser diode 405 and 532 nm as a source, combined with a continuous flow injection technique. The first method depends on the effect of MFA on calcein blue (CLB) fluorescence at 405 nm. Another method is a study of rhodamine-6G (Rh-6G) fluorescence after adding MFA, and recording at 532 nm. Optimum parameters as fluorescent dye concentration, basic medium, flow rate, sample volume, purge time, and delay coil have been investigated. The dynamic range of MFA was 0.2 to 2 mmol.L-1; 0.5 to 2.3 mmol.L-1 with linearity percentage (% r2) 98.92 and 99.83%, for Rh-6G and CLB, respectively. Limit of detection at a minimum concentration in calibration curve 189.34 and 199.89 ng/sample, for Rh-6G and CLB, respectively. The comparison of developed methods with the classical method (UV-vis spectrophotometry) was achieved. The proposed methods were successfully applied for the determination of MFA in the pharmaceutical samples and can be used as an alternative method.

Scopus Crossref
View Publication
Publication Date
Sun Dec 01 2013
Journal Name
Baghdad Science Journal
Construction of New Ion Selective Electrodes for Determination Fe(III) and Their Application in Pharmaceutical samples
...Show More Authors

Liquid membrane electrodes for the determination iron(III) were constructed based on chloramphenicol sodium succinate and iron(III) CPSS-Fe(III) as ion pair complex, with four plasticizers Di-butyl phosphate (DBP); Di-butyl phthalate (DBPH); Di-octyl phthalate (DOP); Tri-butyl phosphate (TBP); in PVC matrix . These electrodes give Nernstian and sub-Nernstian slopes (19.79, 24.60, 16.01 and 13.82mV/decade) and linear ranges from (1x10-5-1x10-2 M, 1x10-5-1x10-2 M, 1x10-6-1x10-2 M and 1x10-5-1x10-2 M) respectively. The best electrode was based on DBP plasticizer which gave a slope 19.79 mV/decade, correlation coefficient 0.9999, detection limit of 9×10-6 M, lifetime 37 day displayed good stability and reproducibility and used to determine

... Show More
View Publication Preview PDF
Crossref
Publication Date
Sun Jun 07 2015
Journal Name
Baghdad Science Journal
Flow Injection Spectrophotometric Determination of Thymol using 4-Aminoantipyrine and Copper(II) Nitrate
...Show More Authors

Flow-injection (FI) spectrophotometric method has been developed for the analysis of thymol in pharmaceutical preparations. The method is based on organic coupling reaction between thymol and 4-amino antipyrine in the presence of alkaline medium to form an intense stable red color complex with copper nitrate that has a maximum absorption at 490 nm. Optimum conditions for determination of the drug was investigated .The calibration graph was linear over the range of 5-500 µg.ml-1 of thymol . The limit of detection (LOD) and limit of quantification (LOQ) were 1.81 ?g mL-1 and 3.60 ?g mL-1 respectively .The proposed method was applied satisfactorily to the determination of thymol in mouth wash preparations. The procedure is characterized by

... Show More
View Publication Preview PDF
Crossref (1)
Crossref
Publication Date
Mon Apr 28 2025
Journal Name
Chemical Papers
New chemiluminometric method for the determination of azithromycin in a continuous flow injection analysis system
...Show More Authors

A novel analytical method is developed for the determination of azithromycin. The method utilizes continuous flow injection analysis to enhance the chemiluminescence system of luminol, H2O2, and Cr(III). The method demonstrated a linear dynamic range of 0.001–100 mmol L-1 with a high correlation coefficient (r) of 0.9978, and 0.001–150 mmol L-1 with a correlation coefficient (r) of 0.9769 for the chemiluminescence emission versus azithromycin concentration. The limit of detection (L.O.D.) of the method was found to be 18.725 ng.50 µL−1 based on the stepwise dilution method for the lowest concentration within the linear dynamic range of the calibration graph. The relative standard deviation (R.S.D. %) for n = 6 was less than 1.2%

... Show More
View Publication
Crossref
Publication Date
Mon Apr 03 2023
Journal Name
Chemistryselect
Development and Validation of Spectrophotometric Methods for the Quantitative Determination of Doxycycline Hyclate in Pure Form and Pharmaceutical Formulations Using Flow‐Injection and Batch procedures: A Comparative Study
...Show More Authors
Abstract<p>Doxycycline hyclate is an antibiotic drug with a broad‐spectrum activity against a variety of gram‐positive and gram‐negative bacteria and is frequently used as a pharmacological agent and as an effector molecule in an inducible gene expression system. A sensitive, reliable and fast spectrophotometric method for the determination of doxycycline hyclate in pure and pharmaceutical formulations has been developed using flow injection analysis (FIA) and batch procedures. The proposed method is based on the reaction between the chromogenic reagent (V<sup>4+</sup>) and doxycycline hyclate in a neutral medium, resulting in the formation of a yellow compound that shows maximum absorbance at 3</p> ... Show More
View Publication
Scopus (5)
Crossref (3)
Scopus Clarivate Crossref
Publication Date
Sun Apr 29 2012
Journal Name
Journal Of Karbala University
Optimization Quantitative Determination of Cimetidine in Pharmaceutical Preparations via Bromothymol Blue Using Central Composite Design
...Show More Authors

Preview PDF
Publication Date
Sun Sep 01 2013
Journal Name
Baghdad Science Journal
High performance liquid chromatographic method for the determination of guaifenesin in pharmaceutical syrups and in environmental samples
...Show More Authors

A simple, precise, rapid, and accurate reversed – phase high performance liquid chromatographic method has been developed for the determination of guaifenesin in pure from pharmaceutical formulations.andindustrial effluent. Chromatography was carried out on supelco L7 reversed- phase column (25cm × 4.6mm), 5 microns, using a mixture of methanol –acetonitrile-water: (80: 10 :10 v/v/v) as a mobile phase at a flow rate of 1.0 ml.min-1. Detection was performed at 254nm at ambient temperature. The retention time for guaifenesin was found 2.4 minutes. The calibration curve was linear (r= 0.9998) over a concentration range from 0.08 to 0.8mg/ml. Limit of detection (LOD) and limit of quantification ( LOQ) were found 6µg/ml and 18µg/ml res

... Show More
View Publication Preview PDF
Crossref
Publication Date
Fri Jan 01 2021
Journal Name
Annals Of Tropical Medicine & Public Health
Newly developed method for determination of methyldopa drug using ammonium ceric (IV) nitrate by continuous flow injection analysis via homemade NAG-ADF-300-2 analyser
...Show More Authors

A new, simple, sensitive and fast developed method was used for the determination of methyldopa in pure and pharmaceutical formulations by using continuous flow injection analysis. This method is based on formation a burgundy color complex between methyldopa andammonium ceric (IV) nitrate in aqueous medium using long distance chasing photometer NAG-ADF-300-2. The linear range for calibration graph was 0.05-8.3 mmol/L for cell A and 0.1-8.5 mmol/L for cell B, and LOD 952.8000 ng /200 µL for cell A and 3.3348 µg /200 µL for cell B respectively with correlation coefficient (r) 0.9994 for cell A and 0.9991 for cell B, RSD % was lower than 1 % for n=8. The results were compared with classical method UV-Spectrophotometric at λ max=280 n

... Show More
Crossref (2)
Crossref
Publication Date
Fri Nov 20 2015
Journal Name
Iraqi Journal Of Science
New Approach for Determination of Oxonium Ion in Inorganic Acids by Quenched Fluorescence of Analytically Interested Species via the Liberated Bromine Using Continuous Flow Injection Laser Diode Fluorimeter Analyser
...Show More Authors

A new method is characterized by simplicity, accuracy and speed for determination of Oxonuim ion in ionisable inorganic acid such as hydrochloric (0.1 - 10) ,Sulphuric ( 0.1 - 6 ),nitric ( 0.1 - 10 ), perchloric ( 0.1 - 7 ), acetic (0.1 - 100 ) and phosphoric ( 0.1 - 30 ) ( mMol.L-1 )acids. By continuous flow injection analysis. The proposed method was based on generation of bromine from the Bro-3-Br-- H3O+. Bromine reacts with fluorescein to quenches the fluorescence . A sample volume no.1 (31μl) and no.2 (35μl) were used with flow rate of 0.95 mL.min-1 using H2O line no.1as carrier stream and 1.3 mL.min-1 using fluorescein sodium salt line no.2. Linear regression of the concentration ( mMol.L-1 ) Vs quenched fluorescence gives a correla

... Show More
View Publication
Publication Date
Tue Jan 01 2019
Journal Name
Research Journal Of Pharmacy And Technology
New Mode Semi-Automated Turbidimetric Determination of Mefenamic Acid by Ayah 6SX1-ST-2D Solar cell-CFI Analyser
...Show More Authors

View Publication
Scopus (1)
Crossref (1)
Scopus Crossref
Publication Date
Sun Dec 07 2008
Journal Name
Baghdad Science Journal
New Method for Determination of Molybdenum(VI) through Flow Injection Analysis Via the Consumption of Liberated Oxygen from Reaction System Ammonia-Hydrogen peroxide-Molybdenum(VI) in the Presence of Pyrogallol
...Show More Authors

This piece of research work aims to study one of the most difficult reaction and determination due to continuous and rapid variation of reaction products and the reactants. As molybdenum (VI) aid in the decomposition of hydrogen peroxide in alkaline medium of ammomia, thus means a continuous liberation of oxygen which cuases and in a continuous manner a distraction in the measurement process. On this basis pyrogallol was used to absorbe all liberated oxygen and the result is an a clean undisturbed signals. Molybdenum (VI) was determined in the range of 4-100 ?g.ml-1 with percentage linearity of 99.8% or (4-300 ?g.ml-1 with 94.4%) while L.O.D. was 3.5 ?g.ml-1. Interferring ions (cations and anions) were studied and their main effect was red

... Show More