Preferred Language
Articles
/
ijs-7853
CFIA-ISNAG fluorimeter for the Determination of Bromhexine-HCl in Drugs via the Measurement Scattered Light at ± 90˚
...Show More Authors

A precise, simple, and accurate continuous flow injection technique was used for the instantaneous estimation of bromhexine hydrochloride (BH-HCl) in tablet dosage form. The chemical and physical parameters of the reaction of BH-HCl with tetraphenylborate to produce a yellowish-white precipitate were determined using an ISNAG fluorimeter analyzer and diverging light at 90°. The calibration curve for BH-HCl was linear with correlation coefficients of 0.9994 and linearity percentage r2% = 99.87 over a concentration range of 0.01-20 mmol/L, L.O.D = 0.3610 µg/125µL (0.007 mmol/L), and RSD% less than 0.3% for 3 and 13 mmol/L (five replicates). This approach was efficiently used to estimate the levels of BH-HCl in two distinct pharmaceutical companies. When the newly developed method was compared with the turbidimetric method using a t-test and standard addition, it was discovered that there had not been any significant difference between the two approaches at the 95% level of confidence.

Scopus Crossref
View Publication Preview PDF
Quick Preview PDF
Publication Date
Sun Mar 04 2018
Journal Name
Iraqi Journal Of Science
Assessment of ISNAG fluorimeter (Total fluorescence measurements at+ 90° & - 90° using four solar cell on each side for 100mm distance at 2mm path length) with well-known fluorescent molecules via CFIA
...Show More Authors

Two well-known fluorescent molecules: fluorescein sodium salt (FSS) and 2,7-dichloro fluorescein (DCF) were tried to prove the efficiency, trustability and repeatability of ISNAG fluorimeter by using discrete and continuous flow injection analysis modes.A linear range of 0.002-1 mmol/L for FSS and 0.003-0.7 mmol/L was for DCF, with LOD 0.0018 mmol/L and 0.002 mmol/L for FSS and DCF respectively, were obtained for discrete mode of analysis. While the continuous mode gave a linear range of 0.002-0.7 mmol/L and 0.003-0.5 mmol/L for FSS and DCF respectively, the LOD were 0.0016mmol/L and 0.0018 mmol/L for FSS and DCF respectively. The results were compared with classical method at variable λex for both fluorescent molecules at 95

... Show More
View Publication Preview PDF
Publication Date
Sun Jun 02 2019
Journal Name
Baghdad Science Journal
New Fluorometric Method for the Determination of Ketotifen Fumarate Using Continuous Flow Injection Analysis via ISNAG-fluorimeter
...Show More Authors

        A newly developed analytical method was conducted for the determination of Ketotifen fumarate (KTF) in pharmaceuticals drugs via quenching of continuous fluorescence of 9(10H)-Acridone (ACD). The method was applied using flow injection system of a new homemade ISNAG fluorimeter with fluorescence measurements at ± 90◦ via 2×4 solar cell. The calibration graph was linear in the range of 1-45 mmol/L, with correlation coefficient r = 0.9762 and the limit of detection 29.785 µg/sample from the stepwise dilution for the minimum concentration in the linear dynamic ranged of the calibration graph. The method was successfully applied to the determination of Ketotifen fumarate in two different pharma

... Show More
View Publication Preview PDF
Scopus (2)
Crossref (1)
Scopus Clarivate Crossref
Publication Date
Fri Dec 08 2023
Journal Name
Iraqi Journal Of Science
Determination of Atenolol in pharmaceutical formulations by continuous flow injection analysis via turbidimetric (T180o) and scattered light effect at two opposite position (2N90o ) using Ayah 4SW-3D-T180 -2N90 -Solar - CFI
...Show More Authors

A simple, rapid and sensitive method for the analysis of Atenolol in pure and pharmaceutical preparation as an alternative analytical procedure were developed by continuous flow injection analysis via turbidimetric (T180o) and scattered light effect at two opposite position (2N90o). The method is based upon the formation of white precipitate for the ion pair compound by phosphomolybidic acid with Atenolol in aqueous medium. The precipitate is measured via the attenuation of incident light and scattering of the incident light in two opposite direction namely +90o and -90o angle were measured. Chemical and physical parameters were investigated. The linearity of Atenolol is ranged from (0.1-11) mmol.L-1, with correlation coefficient r=0.993

... Show More
View Publication Preview PDF
Publication Date
Thu Apr 28 2022
Journal Name
Iraqi Journal Of Science
Two Techniques (Spectrophotometric and Turbidimetric) for Determination of Ciprofloxacin HCl in Pharmaceutical Drugs with Comparison between the Techniques
...Show More Authors

A new, effective, simple and inexpensive method was developed for determination the amounts of Ciprofloxacin HCl in solutions and in tablets by spectrophotometry and turbidity using sodium tungstate as reagent. Also, in this paper comparison between two techniques was conducted. The reaction between the Ciprofloxacin HCl and reagent in acidic media formed a yellowish white complex. Chemical and physical conditions have been investigated. The Linear range for spectroscopic and turbidimetric techniques were (0.05-1.25) and (0.05-3.0) mmol.L-1 with correlation coefficients (0.9956) and (0.9941) respectively. The limits of detection were 7×10-5 and 5×10-5 mol.L-1. The proposed method is good alternative for determination of Ciprofloxacin H

... Show More
View Publication Preview PDF
Publication Date
Mon May 15 2023
Journal Name
Iraqi Journal Of Science
Determination of hydrogen peroxide in some local pharmaceutical disinfectants by continuous flow injection analysis via turbidimetric (T180o) and scattered light effect at two opposite positions (2N90o ) using Ayah 4SW-3D-T180o -2N90o -Solar - CFI Analyse
...Show More Authors

An indirectly method is used to determine hydrogen peroxide. The method based on oxidation of chromium (III) ion by hydrogen peroxide in basic medium to form chromate ion which react with barium (II) ion to produce a yellow precipitate (BaCrO4). Under the optimum established conditions, the linear range of 0.50-25.00 mmol L-1 along with correlation coefficient (r) of 0.9992, Limit of detection (LOD) 0.68 μg / 100 μL, precision expressed as relative standard deviation for six replication measurements at 5.0 mmol.L-1 H2O2 of less than 2% were obtained for hydrogen peroxide. The developed method was successfully applied for the estimation of H2O2 in three pharmaceuticals preparation of different companies using continuous flow injection o

... Show More
View Publication Preview PDF
Publication Date
Sun Dec 01 2013
Journal Name
Baghdad Science Journal
Flow injection analysis for the photometric determination of promethazine-HCl in pure and pharmaceutical preparation via oxidation by persulphate using Ayah 3SX3-3D solar micro photometer
...Show More Authors

The first flow injection spectrophotometric method is characterized by its speed and sensitivity which have been developed for the determination of promethazine-HCl in pure and pharmaceutical preparation. It is based on the in situ detection of colored cationic radicals formed via oxidation of the drug with sodium persulphate to pinkish-red species and the same species was determined by using homemade Ayah 3SX3-3D solar flow injection photometer. Optimum conditions were obtained by using the high intensive green light emitted diode as a source. Linear dynamic range for the absorbance versus promethazine-HCl concentration was 0-7 mmol.L-1, with the correlation coefficient (r) was 0.9904 while the percentage linearity (r2%) was 98.09%. the L.

... Show More
View Publication Preview PDF
Crossref
Publication Date
Tue Sep 29 2020
Journal Name
Iraqi Journal Of Science
Spectrophotometric Determination of Bromhexine Hydrochloride by Diazotization and Coupling Method in Its Pharmaceutical Preparations
...Show More Authors

A simple, fast, and sensitive spectrophotometric method was suggested for the determination of Bromhexine Hydrochloride (BHH) in its pharmaceutical formulations. The method depends on the diazotization of BHH by sodium nitrite in acidic medium to produce the corresponding diazonium salt. The latter is coupled with phloroglucinol reagent in alkali medium to form a yellow water soluble azo-dye which has a maximum absorption at 405 nm with a molar absorptivity of 2.7×104 l.mol-1.cm-1 and Sandellʼs sensitivity of 0.01517 µg.cm-1. Beerʼs low is obeyed within a concentration range of 0.25-15 µg.mL-1 of BHH. The LOD and LOQ values of the proposed method were 0.087 µg.mL

... Show More
View Publication Preview PDF
Scopus (3)
Scopus Crossref
Publication Date
Sun Sep 02 2012
Journal Name
Baghdad Science Journal
Batch and Flow-Injection Spectrophotometric Determination of Procaine HCl in Pharmaceutical Preparations Via Using Diazotization and Coupling Reaction
...Show More Authors

Simple and sensitive batch and Flow-injection spectrophotometric methods for the determination of Procaine HCl in pure form and in injections were proposed. These methods were based on a diazotization reaction of procaine HCl with sodium nitrite and hydrochloric acid to form diazonium salt, which is coupled with chromatropic acid in alkaline medium to form an intense pink water-soluble dye that is stable and has a maximum absorption at 508 nm. A graphs of absorbance versus concentration show that Beer’s law is obeyed over the concentration range of 1-40 and 5-400 µg.ml-1 of Procaine HCl, with detection limits of 0.874 and 3.75 µg.ml-1 of Procaine HCl for batch and FIA methods respectively. The FIA average sample throughput was 70 h-1. A

... Show More
View Publication Preview PDF
Crossref (1)
Crossref
Publication Date
Sat Jun 03 2023
Journal Name
Iraqi Journal Of Science
New mode for the on-line determination of oxonium ion in different strong acids using CFIA via the use of homemade linear array Ayah 5Sx4-ST- 5D solar CFI analyser
...Show More Authors

New mode for the on-line determination of oxonium ion in different strong acids using CFIA via the use of homemade linear array Ayah 5Sx4-ST- 5D solar CFI analyser

View Publication Preview PDF
Publication Date
Sun Jul 02 2023
Journal Name
Iraqi Journal Of Science
Chemiluminometric- CFIA for the Determination of Cobalt (ll) Ion in Commercial Cobalt-Molybdenum Catalyst (K F124-3E) Used in Desulphurization Processes of Petroleum Products via Multi Gel Beads Reactor.
...Show More Authors

Co(II) ion was determined by a new, accurate, sensitive and rapid method via a
continuous flow injection analysis (CFIA) with a chemiluminescence reaction based on
the oxidation of Luminol which is loaded on poly acrylic acid gel beads by hydrogen
peroxide in presence of Cobalt (II) ion as a chemiluminescence catalyst. Chemical and
physical parameters were investigated to obtain the best conditions. Linear dynamic
range of Cobalt (II) ion was from 0.1-20.0 μg.ml-1 with a correlation coefficient r =
0.9758, limit of detection (L.O.D) 0.2 ng/sample from the step wise dilution of lowest
concentration in the calibration graph with the percentage relative standard deviation for
3 μg.ml-1 Co(ll) solution is 0.8537% (n

... Show More
View Publication Preview PDF