A newly flow injection-turbidimetric method characterized by it is speed and sensitivity has been developed for the determination of Amiloride in pure and pharmaceutical preparations. It is based on the formation of yellowish white precipitate for the Amiloride-phosphomolybidic acid ion pair in aqueous medium. Turbidity was measured by Ayah 6Sx1-T-1D solar cell CFI analyser via the attenuation of incident light from the surfaces precipitated particles at 0-180. The Chemical and physical parameters were investigated. Linear dynamic range for the attenuation of incident light versus Amiloride concentration was of 0.005-10 mmol.L-1, with the correlation coefficient (r) of 0.9986 , while the percentage linearity (r2%) was 99.71%. The L.O.Q was of 23.14 μg/sample, while L.O.D (S/N=3) =5.74 ng/sample from the stepwise dilution for the minimum concentration in the calibration graph. The R.S.D% at 3, 9 mmol.L-1 Amiloride is less than 0.5% (eight replicates) using 50 μL sample volume. Through put was of 40 sample.hr-1. The method was applied successfully for the determination of Amiloride in pharmaceutical preparations. A comparison was made using the standard additions method via the use paired t-test. It showed that there was no significant difference between the quoted value of each individual paired t-test with calculated t-value at 95% confidence interval from developed method. In addition to comparison between two methods and calculate t-value, it was noticed that there was no significant difference between the two methods at =0.05 (95% confidence level).
A simple, rapid and sensitive method for the analysis of Atenolol in pure and pharmaceutical preparation as an alternative analytical procedure were developed by continuous flow injection analysis via turbidimetric (T180o) and scattered light effect at two opposite position (2N90o). The method is based upon the formation of white precipitate for the ion pair compound by phosphomolybidic acid with Atenolol in aqueous medium. The precipitate is measured via the attenuation of incident light and scattering of the incident light in two opposite direction namely +90o and -90o angle were measured. Chemical and physical parameters were investigated. The linearity of Atenolol is ranged from (0.1-11) mmol.L-1, with correlation coefficient r=0.993
... Show MoreAn indirectly method is used to determine hydrogen peroxide. The method based on oxidation of chromium (III) ion by hydrogen peroxide in basic medium to form chromate ion which react with barium (II) ion to produce a yellow precipitate (BaCrO4). Under the optimum established conditions, the linear range of 0.50-25.00 mmol L-1 along with correlation coefficient (r) of 0.9992, Limit of detection (LOD) 0.68 μg / 100 μL, precision expressed as relative standard deviation for six replication measurements at 5.0 mmol.L-1 H2O2 of less than 2% were obtained for hydrogen peroxide. The developed method was successfully applied for the estimation of H2O2 in three pharmaceuticals preparation of different companies using continuous flow injection o
... Show MoreA simple, accurate and sensitive spectrophotometric way is used to determine Bisacodyl in pure and pharmaceutical preparations. The proposed method depends on using 2,4-Dinitrophenylhydrazine as chromogenic reagent . The method was based on the oxidative coupling reaction of Bisacodyl with 2,4-Dinitrophenylhydrazine with Sodium periodate in the presence of sodium hydroxide as alkaline media to form red water soluble dye product , that has a maximum absorption at ?max 522nm . Beer ,s law is obeyed in the concentration of (2.00–20.00) ?g.ml -1 .The molar absorptivity is (6505) L.mol-1.cm-1,a sandall sensitivity of(0.0555) ?g.cm-2), correlation coefficient of (0.9970) , Limitof detection (LOD) (0.0312 ?g.ml-1), limit of Quantitation (LOQ) (
... Show MoreA new way to Systems concentrates have been clarified and that allows a concentration high and analysis to automatically wavelengths of the spectrum of this system analyst of the spectrum and the center is built on Holucram Nafez gives less absorbency with efficient diffraction high when the wavelength (900 nm), which will be useful for Khallaya solar
Two well-known fluorescent molecules: fluorescein sodium salt (FSS) and 2,7-dichloro fluorescein (DCF) were tried to prove the efficiency, trustability and repeatability of ISNAG fluorimeter by using discrete and continuous flow injection analysis modes.A linear range of 0.002-1 mmol/L for FSS and 0.003-0.7 mmol/L was for DCF, with LOD 0.0018 mmol/L and 0.002 mmol/L for FSS and DCF respectively, were obtained for discrete mode of analysis. While the continuous mode gave a linear range of 0.002-0.7 mmol/L and 0.003-0.5 mmol/L for FSS and DCF respectively, the LOD were 0.0016mmol/L and 0.0018 mmol/L for FSS and DCF respectively. The results were compared with classical method at variable λex for both fluorescent molecules at 95
... Show MoreAn accurate and sensitive spectrophotometric method has been developed for the determination of cefotaxime (CEF) in pure and pharmaceutical samples. The suggested method depended on the coupling reaction between diazotized cefotaxime and 3,5-dimethyl phenol (3,5-DMPH) in basic medium to form light orange, water soluble dye, that is stable and has a maximum absorbance at 497nm. The calibration graph was liner over the concentration range (1-70) µg.mL-1 with LOD of 0.750 µg.mL-1 and LOQ of. 2.740 µg. mL-1, sandal sensitivity of 0.0526 µg. cm-2 . molar absorptivity 11328 Lmol-1 cm-1 . The stoichiometry composition was found by Jobs a
... Show MoreTwo simple methods spectrophotometric were suggested for the determination of Cefixime (CFX) in pure form and pharmaceutical preparation. The first method is based without cloud point (CPE) on diazotization of the Cefixime drug by sodium nitrite at 5Cº followed by coupling with ortho nitro phenol in basic medium to form orange colour. The product was stabilized and measured 400 nm. Beer’s law was obeyed in the concentration range of (10-160) μg∙mL-1 Sandell’s sensitivity was 0.0888μg∙cm-1, the detection limit was 0.07896μg∙mL-1, and the limit of Quantitation was 0.085389μg∙mL-1.The second method was cloud point extraction (CPE) with using Trtion X-114 as surfactant. Beer
... Show MoreThe research involved a rapid, automated and highly accurate developed CFIA/MZ technique for estimation of phenylephrine hydrochloride (PHE) in pure, dosage forms and biological sample. This method is based on oxidative coupling reaction of 2,4-dinitrophenylhydrazine (DNPH) with PHE in existence of sodium periodate as oxidizing agent in alkaline medium to form a red colored product at ʎmax )520 nm (. A flow rate of 4.3 mL.min-1 using distilled water as a carrier, the method of FIA proved to be as a sensitive and economic analytical tool for estimation of PHE.
Within the concentration range of 5-300 μg.mL-1, a calibration curve was rectilinear, where the detection limit was 3.252 μg.mL