Preferred Language
Articles
/
4xePeI0BVTCNdQwC-xQ0
NEW APPROACH FOR DETERMINATION OF MEBEVERINE BY QUENCHED FLUORESCENCE OF ANALYTICALLY INTERESTED SPECIES USING CONTINUOUS FLOW INJECTION LASER DIODE FLUORIMETER ANALYSER
...Show More Authors

The proposed method is sensitive, simple , fast for the determination of mebeverine hydrochloride in pure form or in pharmaceutical dosage . Using Homemade instrument fluorimeter continuous flow injection analyser with solid state laser (405 nm) as a source. Where it is based upon the fluorescence of fluorescein sodium salt and quenching effect of fluorescence by mebeverine in aqueous medium. The calibration graph was linear in the concentration range 0.05 to10 mMol.L-1 (r= 0.9629) with relative standard deviation (RSD%) for 1 mMol.L-1mebeverine solution was lower than 3% (n=6). Three pharmaceutical drugs were used as an application for the determination of mebeverine. A comparison was made between the newly developed method of analysis with the quoted value using the standard addition method. It can be noticed that there was no significant differences between the newly developed method and the quoted value by the manufacturers companies. It indicates clearly that the new method can be used for the assessment as well as the method adopted by the manufactures companies that can be described.

View Publication